INTERFERENCE BY HEMOLYSIS, ICTERUS AND LIPEMIA IN ASSAYS ON THE BECKMAN SYNCHRON-CX5 AND METHODS FOR CORRECTION

被引:33
作者
RANDALL, AG
GARCIAWEBB, P
BEILBY, JP
机构
[1] Dept. of Clinical Biochemistry, Queen Elizabeth II Med. Center, Nedlands
关键词
Automated analysis; Bichromatic spectroscopy;
D O I
10.1177/000456329002700411
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
As part of an evaluation of a Synchron CX5 analyser (Beckman Instruments Inc, Brea, USA) we examined a range of tests for interference from haemolysis, bilirubin and lipaemia. Tests investigated were urea, creatinine, urate, total protein, albumin, calcium, total bilirubin, alkaline phosphatase (ALP), aspartate transaminase (AST), λ-glutamyl transferase (GGT) and inorganic phosphate. Two types of interferences were found. One type is found on other analysers and represents analytical difficulties with the measurement of that particular analyte. The other type of interference was a consequence of the bichromatic optical system used on the CX-5. This latter group includes haemoglobin interference in the measurement of total protein and inorganic phosphate, and bilirubin interference with the measurement of total protein, glucose and inorganic phosphate. Lipaemia interfered with total protein, total bilirubin, inorganic phosphate, urate and glucose. Alternative and modified methods are proposed to improve the measurement of total protein, glucose, total bilirubin and inorganic phosphate. The use of the modified methods for glucose, inorganic phosphate and total bilirubin are limited, at this time, by an error in the calculation algorithm used by the analyser for two step or triggered chemistries, and to a lesser extent, by a reduction in sample throughput.
引用
收藏
页码:345 / 352
页数:8
相关论文
共 14 条
[1]   HEMOLYSIS AND PLASMA ALKALINE-PHOSPHATASE ACTIVITY [J].
BALLANTYNE, FC ;
BORLAND, W ;
BIRRELL, RC ;
DAVIDSON, H .
ANNALS OF CLINICAL BIOCHEMISTRY, 1988, 25 :192-193
[2]  
BOYETT JD, 1975, CLIN CHEM, V21, P1542
[3]  
DENEGAAR CP, 1983, CLIN CHIM ACTA, V131, P155
[4]  
GLICK MR, 1987, CLIN CHEM, V33, P1453
[5]   THE EFFECT OF HEMOLYSIS ON THE MEASUREMENT OF PLASMA ALKALINE-PHOSPHATASE ACTIVITY [J].
GROSSET, A ;
KNAPP, ML ;
MAYNE, PD .
ANNALS OF CLINICAL BIOCHEMISTRY, 1987, 24 :513-517
[6]   INVESTIGATIONS INTO NEGATIVE INTERFERENCE BY JAUNDICED PLASMA IN KINETIC JAFFE METHODS FOR PLASMA CREATININE DETERMINATION [J].
KNAPP, ML ;
HADID, O .
ANNALS OF CLINICAL BIOCHEMISTRY, 1987, 24 :85-97
[7]  
MAGUIRE GA, 1985, CLIN CHEM, V31, P2007
[8]  
MORI L, 1978, CLIN CHEM, V24, P1841
[9]  
PEARSON JR, 1976, CLIN CHEM, V22, P559
[10]  
SHULL BC, 1980, CLIN CHEM, V26, P26