INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY WITH AN ENLARGED SAMPLING ORIFICE AND OFFSET ION LENS .1. ION TRAJECTORIES AND DETECTOR PERFORMANCE

被引:29
作者
HU, K [1 ]
CLEMONS, PS [1 ]
HOUK, RS [1 ]
机构
[1] IOWA STATE UNIV SCI & TECHNOL,US DOE,DEPT CHEM,AMES LAB,AMES,IA 50011
关键词
D O I
10.1016/1044-0305(93)85038-Y
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new inductively coupled plasma mass spectrometer (ICP-MS) with four stages of differential pumping is described. The relatively large sampling orifice (1.31-mm dia.) improves signals for metal ions and resists plugging from deposited solids. A new ion lens is described that deflects ions off center and then back on center into the differential pumping orifice; there is no photon stop in the center of the beam. Calculations of ion trajectories using SIMION show that only those ions that leave the skimmer on center are transmitted, whereas most other lenses used in ICP-MS transmit only ions that leave the skimmer off axis. The performance of a Channeltron electron multiplier is compared to that of a Daly detector. Both detectors yield similar sensitivities of approximately 10(6) counts s-1 per ppm and detection limits of approximately 1 pptr. The background with a Channeltron electron multiplier is only 0.4 counts s-1 and is only slightly higher than the dark current count rate. Presumably the offset ion lens used in the present work efficiently screens the detector from photons emitted by the plasma. The background with the Daly detector is 4 counts s-1, which represents a substantial improvement over the background obtained in pervious use of the Daly detector with ICP-MS.
引用
收藏
页码:16 / 27
页数:12
相关论文
共 34 条
[1]  
BEAR BR, 1986, SPECTROCHIM ACTA B, V41, P1089
[2]   INDUCTIVELY COUPLED PLASMA AS AN ION-SOURCE FOR HIGH-RESOLUTION MASS-SPECTROMETRY [J].
BRADSHAW, N ;
HALL, EFH ;
SANDERSON, NE .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (08) :801-803
[3]  
BRADSHAW N, 1989, Patent No. 622
[4]   FUNDAMENTAL-STUDIES OF THE SAMPLING PROCESS IN AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER .2. ION KINETIC-ENERGY MEASUREMENTS [J].
CHAMBERS, DM ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (6-7) :761-784
[5]   MASS-SPECTROMETRIC MEASUREMENT OF IONIZATION TEMPERATURE IN AN INDUCTIVELY COUPLED PLASMA [J].
CRAIN, JS ;
SMITH, FG ;
HOUK, RS .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (03) :249-259
[6]  
DAHL DA, 1990, PC PS2 SIMION VERSIO
[7]  
DALY NR, 1960, REV SCI INSATRUM, V37, P1385
[8]   GAS-DYNAMICS OF THE INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY INTERFACE [J].
DOUGLAS, DJ ;
FRENCH, JB .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (06) :743-747
[9]   STUDY OF SOLIDS DEPOSITION ON INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY SAMPLERS AND SKIMMERS [J].
DOUGLAS, DJ ;
KERR, LA .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (06) :749-752
[10]   ION KINETIC ENERGIES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY (ICP-MS) [J].
FULFORD, JE ;
DOUGLAS, DJ .
APPLIED SPECTROSCOPY, 1986, 40 (07) :971-974