DETERMINATION OF WARFARIN IN DRINKING-WATER BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AFTER SOLID-PHASE EXTRACTION

被引:10
作者
DALBACKE, J
DAHLQUIST, I
PERSSON, C
机构
[1] Analytical and Quality Development Department, Gacell Laboratories AB, S-201 80 Malmö
来源
JOURNAL OF CHROMATOGRAPHY | 1990年 / 507卷
关键词
D O I
10.1016/S0021-9673(01)84215-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method for the determination of 0.1 μg/l of warfarin in drinking water involving concentration by solid-phase extraction is described. A 1000-ml volume of drinking water is aspirated through a solid-phase extraction column and warfarin is eluted to 1.0 ml. The eluate is analysed by reversed-phase high-performance liquid chromatography with UV detection. The detection limit is 0.02 μg/l. Recoveries of greater than 90% were obtained when tap water was spiked with warfarin. © 1990.
引用
收藏
页码:381 / 387
页数:7
相关论文
共 10 条
[1]   NOVEL METHOD FOR THE QUANTITATION OF WARFARIN AND ITS METABOLITES IN PLASMA [J].
FASCO, MJ ;
CASHIN, MJ ;
KAMINSKY, LS .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1979, 2 (04) :565-575
[2]   BIOCHEMICAL APPLICATIONS OF A QUANTITATIVE HIGH-PRESSURE LIQUID-CHROMATOGRAPHIC ASSAY OF WARFARIN AND ITS METABOLITES [J].
FASCO, MJ ;
PIPER, LJ ;
KAMINSKY, LS .
JOURNAL OF CHROMATOGRAPHY, 1977, 131 (JAN21) :365-373
[3]   GLC DETERMINATION OF WARFARIN IN HUMAN-PLASMA [J].
HANNA, S ;
ROSEN, M ;
EISENBERGER, P ;
RASERO, L ;
LACHMAN, L .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1978, 67 (01) :84-86
[4]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION AND FLUORESCENCE DETECTION OF WARFARIN AND ITS METABOLITES BY POST-COLUMN ACID-BASE MANIPULATION [J].
LEE, SH ;
FIELD, LR ;
HOWALD, WN ;
TRAGER, WF .
ANALYTICAL CHEMISTRY, 1981, 53 (03) :467-471
[5]   REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF WARFARIN FROM BIOLOGICAL-FLUIDS IN THE LOW NANOGRAM RANGE [J].
STEYN, JM ;
VANDERMERWE, HM ;
DEKOCK, MJ .
JOURNAL OF CHROMATOGRAPHY, 1986, 378 (01) :254-260
[6]   RAPID, RELIABLE AND SENSITIVE ASSAY FOR WARFARIN USING NORMAL-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
TASKER, RAR ;
NAKATSU, K .
JOURNAL OF CHROMATOGRAPHY, 1982, 228 (MAR) :346-349
[7]   SENSITIVE MEASUREMENTS OF WARFARIN BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION [J].
WANG, J ;
BONAKDAR, M .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 415 (02) :432-437
[8]   ANALYSIS OF WARFARIN IN PLASMA BY HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
WONG, LT ;
SOLOMONRAJ, G ;
THOMAS, BH .
JOURNAL OF CHROMATOGRAPHY, 1977, 135 (01) :149-154
[9]  
1980, EEC DRINKING WATER G
[10]  
1985, US PHARMACOPEIA, P1228