BACKSCATTER INTO THE BEAM MONITOR CHAMBER - IMPLICATIONS FOR DOSIMETRY OF ASYMMETRIC COLLIMATORS

被引:45
作者
DUZENLI, C
MCCLEAN, B
FIELD, C
机构
关键词
BEAM MONITOR CHAMBER (BMC); ASYMMETRIC COLLIMATORS; BACKSCATTER;
D O I
10.1118/1.597077
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Backscatter from the asymmetric collimators of a linac into the beam monitor chamber (BMC) has been investigated for two accelerators having different BMC configurations. The effect has been quantified as a function of field size and collimator jaw position for 6 and 18 MV beams. The results indicate a maximum 2.5% (6 MV) and 4% (18 MV) decrease in output in one case and a negligible effect in the other case. The experiments indicate that the difference can be attributed to the different construction of the BMC's for the two accelerators.