Evaluation of the iPhone with an acrylic sleeve versus the Scoliometer for rib hump measurement in scoliosis

被引:32
作者
Izatt, Maree T. [1 ,2 ]
Bateman, Gary R. [3 ]
Adam, Clayton J. [1 ,2 ]
机构
[1] Queensland Univ Technol, Paediatr Spine Res Grp, South Brisbane, Qld 4101, Australia
[2] Mater Childrens Hosp, Mater Hlth Serv, Level 2, South Brisbane, Qld 4101, Australia
[3] GB Orthopaed Australia, South Brisbane, Qld 4101, Australia
来源
SCOLIOSIS AND SPINAL DISORDERS | 2012年 / 7卷
关键词
Scoliosis; Rib hump; Rib hump measurement; Scoliometer; iPhone; Measurement variability; Smartphone; Inter-observer variability; Intra-observer variability;
D O I
10.1186/1748-7161-7-14
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Vertebral rotation found in structural scoliosis contributes to trunkal asymmetry which is commonly measured with a simple Scoliometer device on a patient's thorax in the forward flexed position. The new generation of mobile 'smartphones' have an integrated accelerometer, making accurate angle measurement possible, which provides a potentially useful clinical tool for assessing rib hump deformity. This study aimed to compare rib hump angle measurements performed using a Smartphone and traditional Scoliometer on a set of plaster torsos representing the range of torsional deformities seen in clinical practice. Methods: Nine observers measured the rib hump found on eight plaster torsos moulded from scoliosis patients with both a Scoliometer and an Apple iPhone on separate occasions. Each observer repeated the measurements at least a week after the original measurements, and were blinded to previous results. Intra-observer reliability and inter-observer reliability were analysed using the method of Bland and Altman and 95% confidence intervals were calculated. The Intra-Class Correlation Coefficients (ICC) were calculated for repeated measurements of each of the eight plaster torso moulds by the nine observers. Results: Mean absolute difference between pairs of iPhone/Scoliometer measurements was 2.1 degrees, with a small (1 degrees) bias toward higher rib hump angles with the iPhone. 95% confidence intervals for intra-observer variability were +/- 1.8 degrees (Scoliometer) and +/- 3.2 degrees (iPhone). 95% confidence intervals for inter-observer variability were +/- 4.9 degrees (iPhone) and +/- 3.8 degrees (Scoliometer). The measurement errors and confidence intervals found were similar to or better than the range of previously published thoracic rib hump measurement studies. Conclusions: The iPhone is a clinically equivalent rib hump measurement tool to the Scoliometer in spinal deformity patients. The novel use of plaster torsos as rib hump models avoids the variables of patient fatigue and discomfort, inconsistent positioning and deformity progression using human subjects in a single or multiple measurement sessions.
引用
收藏
页数:7
相关论文
共 14 条
[1]  
Adams W, 1865, LECT PATHOLOGY TREAT
[2]   VALIDITY AND RELIABILITY TESTING OF THE SCOLIOMETER [J].
AMENDT, LE ;
AUSEELLIAS, KL ;
EYBERS, JL ;
WADSWORTH, CT ;
NIELSEN, DH ;
WEINSTEIN, SL .
PHYSICAL THERAPY, 1990, 70 (02) :108-117
[3]   Applying the right statistics: analyses of measurement studies [J].
Bland, JM ;
Altman, DG .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2003, 22 (01) :85-93
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Inter and intra-rater reliability of the scoliometer [J].
Bonagamba, Guilherme H. ;
Coelho, Daniel M. ;
de Oliveira, Anamaria S. .
BRAZILIAN JOURNAL OF PHYSICAL THERAPY, 2010, 14 (05) :432-437
[6]   AN OBJECTIVE CRITERION FOR SCOLIOSIS SCREENING [J].
BUNNELL, WP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (09) :1381-1387
[7]   MEASUREMENT OF SCOLIOSIS AND KYPHOSIS RADIOGRAPHS - INTRAOBSERVER AND INTEROBSERVER VARIATION [J].
CARMAN, DL ;
BROWNE, RH ;
BIRCH, JG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (03) :328-333
[8]   A study of the diagnostic accuracy and reliability of the scoliometer and Adam's forward bend test [J].
Côté, P ;
Kreitz, BG ;
Cassidy, JD ;
Dzus, AK ;
Martel, J .
SPINE, 1998, 23 (07) :796-802
[9]   Reliability assessment and approaches to determining agreement between measurements: Classic methods paper [J].
Griffiths, Peter ;
Murrells, Trevor .
INTERNATIONAL JOURNAL OF NURSING STUDIES, 2010, 47 (08) :937-938
[10]   The effect of growth on the correlation between the spinal and rib cage deformity: implications on idiopathic scoliosis pathogenesis [J].
Grivas, Theodoros B. ;
Vasiliadis, Elias S. ;
Mihas, Constantinos ;
Savvidou, Olga .
SCOLIOSIS AND SPINAL DISORDERS, 2007, 2