THE NIST PRIMARY RN-222 MEASUREMENT SYSTEM

被引:26
作者
COLLE, R
HUTCHINSON, JMR
UNTERWEGER, MP
机构
[1] Natl Inst of Standards and, Technology, , MD
关键词
D O I
10.6028/jres.095.018
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Within the United States, the national standard for radon measurements is embodied in a primary radon measurement system that has been maintained for over 50 years to accurately measure radon (222Rn) against international and national radium (226Ra) standards. In turn, all of the radon measurements made at the National Institute of Standards and Technology (NIST) and the radon transfer calibration standards and calibration services provided by NIST are directly related to this national radon standard. This primary radon measurement system consists of pulse ionization chambers and ancillary gas handling and gas purification equipment. The system is currently undergoing a significant upgrading and expansion which will replace the extant outdated system.
引用
收藏
页码:155 / 165
页数:11
相关论文
共 19 条
[1]  
BLANCHARD RL, 1964, USDHEW PHS999RH9 PUB
[2]  
COLLE R, 1981, T AM NUCL SOC, V39, P84
[3]  
COLLE R, 1985, 30TH ANN M HLTH PHYS, P185
[4]  
COLLE R, 1981, NBS1139 TECH NOTE
[5]  
COLLE R, 1988, 1ST P INT SUMM SCH G, P79
[6]  
CURTISS IF, 1943, J RES NATL BUR STAND, V31, P181
[7]   COMPARISON OF 4 NATIONAL RADIUM STANDARDS .1. EXPERIMENTAL PROCEDURES AND RESULTS .2. STATISTICAL PROCEDURES AND SURVEY [J].
DAVENPORT, TI ;
MANN, WB ;
MCCRAVEN, CC ;
SMITH, CC ;
CONNOR, WS ;
YOUDEN, WJ .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1954, 53 (05) :267-275
[8]   FACTORS AFFECTING OPERATION OF APPARATUS FOR COUNTING ALPHA PARTICLES IN AN ION COUNTING CHAMBER [J].
DAVIS, FJ .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1947, 39 (06) :545-549
[9]   Apparatus for the determination of minute quantities of radium, radon and thoron in solids, liquids and gases [J].
Evans, RD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1935, 6 (04) :99-112
[10]  
Evans RD, 1937, AM J ROENTGENOL RADI, V37, P368