INSPIRATORY VALVE MALFUNCTION IN A CIRCLE SYSTEM - PITFALLS IN CAPNOGRAPHY

被引:11
作者
KUMAR, AY
BHAVANISHANKAR, K
MOSELEY, HSL
DELPH, Y
机构
[1] Department of Anaesthesia, Queen Elizabeth Hospital, University of West Indies, Faculty of Medical Sciences
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1992年 / 39卷 / 09期
关键词
EQUIPMENT; ANESTHESIA CIRCUITS; MEASUREMENT TECHNIQUES; CAPNOMETRY; MONITORING; CARBON DIOXIDE;
D O I
10.1007/BF03008353
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Capnography is a useful technique in monitoring the integrity of anaesthetic equipment such as the malfunctioning of unidirectional valves in circle system. However, the lack of a precise mechanism in existing capnographs to identify the start of inspiration and the beginning of expiration in the capnograms, makes the analysis of the carbon dioxide waveforms during inspiration difficult and thus results in inaccurate assessment Of rebreathing. We report a case where, during the malfunction of the inspiratory unidirectional valve in the circle system, the capnograph failed to detect the presence of substantial rebreathing. Critical analysis of the capnogram recorded during the malfunction revealed that there was substantial rebreathing which was underestimated by the capnograph as it reports only the lowest CO2 concentration rebreathed during inspiration in such abnormal situations.
引用
收藏
页码:997 / 999
页数:3
相关论文
共 5 条
[1]   CAPNOGRAPHIC DETECTION OF ANESTHESIA CIRCLE VALVE MALFUNCTIONS [J].
BERMAN, LS ;
PYLES, ST .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1988, 35 (05) :473-475
[2]  
PODRAZA AG, 1991, ANESTHESIOLOGY, V75, pA422
[3]  
PYLES ST, 1984, ANESTH ANALG, V63, P536
[4]  
VANGENDERINGEN HR, 1987, J CLIN MONITOR, V3, P194
[5]   RESPIRATORY MONITORING OF CARBON-DIOXIDE AND OXYGEN - A 10-YEAR PERSPECTIVE [J].
WEINGARTEN, M .
JOURNAL OF CLINICAL MONITORING, 1990, 6 (03) :217-225