PROPERTIES OF LIQUID IONIZATION CHAMBERS AT LDR BRACHYTHERAPY DOSE-RATES

被引:11
作者
JOHANSSON, B
WICKMAN, G
HOLMSTROM, T
机构
[1] Dept. of Radiat. Phys., Umea Univ.
关键词
D O I
10.1088/0031-9155/40/4/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Properties such as sensitivity, general recombination and reproducibility of liquid-filled parallel-plate ionization chambers for dosimetry in low-dose-rate brachytherapy radiation fields have been evaluated. Two different dielectric liquids, isooctane (C(8)H(1)8) and tetramethylsilane (Si(CH3)(4)), have been used as sensitive media in chambers having a coin-shaped sensitive volume of 3 mm in diameter and 1 mm thickness. An electric field strength of 300 kV m(-1) was found to be optimal with respect to sensitivity, leakage current and general recombination. At absorbed dose rates from 0.1-100 mGy min(-1) the ionization charge measurements at an irradiation time of 1 min showed a reproducibility better than 1%, and a general recombination not exceeding 0.5%. The calibration-absorbed dose to water against ionization charge at a Co-60 reference source-did not show any significant change over an observation time of one year for any of the chambers.
引用
收藏
页码:575 / 587
页数:13
相关论文
共 20 条
[1]  
BRAHME A, 1976, Medical Physics (Woodbury), V3, P95, DOI 10.1118/1.594210
[2]   A NONINVASIVE METHOD FOR FRACTIONATED STEREOTACTIC IRRADIATION OF BRAIN-TUMORS WITH LINEAR-ACCELERATOR [J].
HARIZ, MI ;
HENRIKSSON, R ;
LOFROTH, PO ;
LAITINEN, LV ;
SATERBORG, NE .
RADIOTHERAPY AND ONCOLOGY, 1990, 17 (01) :57-72
[3]   ELECTRON DEPTH ABSORBED DOSES FOR SMALL PHANTOM DEPTHS - COMPARISON BETWEEN DIFFERENT ACCELERATORS [J].
HULTEN, G ;
SVENSSON, H .
ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1975, 14 (06) :537-544
[4]  
JAFFE MG, 1906, J PHYSIQUE THEOR, V5, P263
[5]   LIQUID IONIZATION-CHAMBER FOR ABSORBED DOSE DETERMINATIONS IN PHOTON AND ELECTRON-BEAMS [J].
JOHANSSON, KA ;
SVENSSON, H .
ACTA RADIOLOGICA ONCOLOGY, 1982, 21 (05) :359-367
[6]   EXPERIMENTAL STUDY OF DOSE DISTRIBUTION IN WATER AROUND CA-137 TUBES USED IN BRACHYTHERAPY [J].
KLEVENHAGEN, SC .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1073-1082
[7]   EFFECT OF FINITE SIZE OF IONIZATION CHAMBERS ON MEASUREMENTS OF SMALL PHOTON SOURCES [J].
KONDO, S ;
RANDOLPH, ML .
RADIATION RESEARCH, 1960, 13 (01) :37-60
[8]   MEASUREMENT OF DOSE DISTRIBUTION AROUND FLETCHER-SUIT-DELCOS COLPOSTATS USING A THERADOS RADIATION-FIELD ANALYZER (RFA-3) [J].
LING, CC ;
SPIRO, IJ ;
KUBIATOWICZ, DO ;
GERGEN, J ;
PEKSENS, RK ;
BENNETT, JD ;
GAGNON, WF .
MEDICAL PHYSICS, 1984, 11 (03) :326-330
[9]   ABSORBED DOSE IN WATER - COMPARISON OF SEVERAL METHODS USING A LIQUID IONIZATION-CHAMBER [J].
MATTSSON, O ;
SVENSSON, H ;
WICKMAN, G ;
DOMEN, SR ;
PRUITT, JS ;
LOEVINGER, R .
ACTA ONCOLOGICA, 1990, 29 (02) :235-240
[10]   DOSIMETRIC CHARACTERISTICS OF A DOUBLE-WALL I-125 SOURCE FOR INTERSTITIAL BRACHYTHERAPY [J].
NATH, R ;
MELILLO, A .
MEDICAL PHYSICS, 1993, 20 (05) :1475-1483