摘要
The suitability of a natural diamond detector with a special contact system for the measurement of relative dose distributions in selected radiotherapy applications was studied. The performance of the diamond detector was analysed by comparison with measurements using an ionization chamber and a silicon diode detector. The good stability, high response and good spatial resolution of the diamond detector were demonstrated by relative measurements in high-energy photon and electron beams and wound the Cs-137 source of an afterloading device. The application of the diamond detectors for relative as well as absolute dose measurements of beta ray ophthalmic applicators was also highlighted. It is concluded that the diamond with the special contact system is a suitable detector for relative dose measurements in a wide variety of applications, especially when high response and high spatial resolution are required.
引用
收藏
页码:173 / 184
页数:12
相关论文
共 17 条
[1]  
BURGEMEISTER EA, 1983, 8TH P S MICR JUL
[2]  
DAVELAAR J, 1989, 1989 P TUM EY ESS
[3]   SYNTHETIC DIAMONDS AS IONIZATION-CHAMBER RADIATION DETECTORS IN BIOLOGICAL ENVIRONMENTS [J].
KEDDY, RJ ;
NAM, TL ;
BURNS, RC .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (06) :751-759
[4]  
KHRUNOV VS, 1990, RADIAT PROT DOSIM, V33, P155
[5]  
KONOROVA EA, 1971, SOV PHYS SEMICOND+, V4, P1600
[6]   DIAMOND DOSIMETER FOR X-RAY AND GAMMA-RADIATION [J].
KOZLOV, SF ;
KONOROVA, EA ;
KUZNETSOV, YA ;
SALIKOV, YA ;
REDKO, VI ;
GRINBERG, VR ;
MEILMAN, ML .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (01) :235-237
[7]   PREPARATION AND CHARACTERISTICS OF NATURAL DIAMOND NUCLEAR RADIATION DETECTORS [J].
KOZLOV, SF ;
STUCK, R ;
HAGEALI, M ;
SIFFERT, P .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1975, NS22 (01) :160-170
[8]   DOSIMETRY OF RU-106 EYE APPLICATORS WITH A P-TYPE SILICON DETECTOR [J].
LAX, I .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :963-972
[9]  
Nordic Association of Clinical Physics, 1980, ACTA RADIOL ONCOL, V19, P55
[10]  
PITKANEN M, 1985, 14 P INT C MED BIOL