USE OF FOCUSED MICROWAVES FOR EXPEDITIVE SHORTENING OF SAMPLE PRETREATMENT - DIGESTION AND REDUCTION PROCEDURES PRIOR TO SELENIUM SPECIATION AS SELENIUM(IV) OR SELENIUM(VI)

被引:35
作者
BRYCE, DW
IZQUIERDO, A
DECASTRO, MDL
机构
关键词
SELENIUM; SPECIATION; MICROWAVE DIGESTION; SAMPLE TREATMENT;
D O I
10.1039/an9952002171
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Digestion programmes for different types of samples (milk, sausage and lyophilized pig kidney certified reference material) have been developed using a focused microwave digestor prior to the determination of selenium. The digestion step does not alter the oxidation state of the analyte; thus, the speciation of the target analyte is also achieved with the help of the microwave digestor, which facilitates a subsequent on-line reduction step in a flow injection (FI) manifold before determination by FI cathodic stripping voltammetry. Excellent recoveries for the spiked and certified values were obtained using digestion programmes which varied from 20-45 min. The on-line microwave procedure for the reduction of Se-VI affords good sensitivity and linearity (3-120 mu g l(-1), r = 0.994) of the over-all determination method, and dramatically reduces the time necessary for Se-VI reduction.
引用
收藏
页码:2171 / 2174
页数:4
相关论文
共 22 条
[1]  
BITSCH R, 1994, LABOR PRAXIS, V18, P76
[2]  
BRUCE DW, 1995, FRESEN J ANAL CHEM, V351, P433
[3]   DETERMINATION OF ZINC AND CADMIUM IN SMALL AMOUNTS OF BIOLOGICAL TISSUES BY MICROWAVE-ASSISTED DIGESTION AND FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY [J].
BURGUERA, M ;
BURGUERA, JL ;
ALARCON, OM .
ANALYTICA CHIMICA ACTA, 1988, 214 (1-2) :421-427
[4]   DETERMINATION OF LEAD IN WINE, OTHER BEVERAGES AND FRUIT SLURRIES BY FLOW-INJECTION HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY WITH ONLINE MICROWAVE DIGESTION [J].
CABRERA, C ;
MADRID, Y ;
CAMARA, C .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (12) :1423-1426
[5]   DETERMINATION OF SELENIUM IN BLOOD AND PLANT MATERIAL BY HYDRIDE GENERATION AND ATOMIC-ABSORPTION SPECTROSCOPY [J].
CLINTON, OE .
ANALYST, 1977, 102 (1212) :187-192
[6]   ONLINE MICROWAVE-ASSISTED DIGESTION OF SOLID SAMPLES FOR THEIR FLAME ATOMIC SPECTROMETRIC ANALYSIS [J].
DELAGUARDIA, M ;
CARBONELL, V ;
MORALESRUBIO, A ;
SALVADOR, A .
TALANTA, 1993, 40 (11) :1609-1617
[7]   COMPARISON OF 2 DIGESTION METHODS FOR THE DETERMINATION OF SELENIUM IN BIOLOGICAL SAMPLES [J].
DUCROS, V ;
RUFFIEUX, D ;
BELIN, N ;
FAVIER, A .
ANALYST, 1994, 119 (08) :1715-1717
[8]  
FERNANDEZ MGC, 1995, FRESEN J ANAL CHEM, V351, P438
[9]  
FIDLER R, 1994, ANAL EUROPA, V2, P39
[10]  
Gao Q., 1994, Fenxi Huaxue = Chinese Journal of Analytical Chemistry, V22, P418