COMPARISON OF SCREW LOOSENING, ROTATION, AND DEFLECTION AMONG 3 IMPLANT DESIGNS

被引:111
作者
DIXON, DL [1 ]
BREEDING, LC [1 ]
SADLER, JP [1 ]
MCKAY, ML [1 ]
机构
[1] UNIV KENTUCKY,COLL DENT,DEPT ORAL HLTH PRACTICE,LEXINGTON,KY
关键词
D O I
10.1016/S0022-3913(05)80134-9
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
A common problem associated with single tooth implant restorations is abutment screw loosening. Manufacturers of implants have attempted to overcome this problem by incorporating antirotational design characteristics into their systems. Micromovement and torque levels required to loosen abutment screws for straight and angled antirotational screw-retained abutment/implant combinations from three different manufacturers were examined in this in vitro investigation. A custom-built machine was used and each sample was subjected to compressive horizontal reciprocal movements over a 25-degree incline for a simulated 1-month period. Data were generated that showed movements of the crown/abutment complex during force application. The amount of torque necessary to loosen the abutment screws before and after testing was also recorded and compared for each system. The results indicated no significant differences (p < 0.05) among all the straight and angled abutments for the variables studied.
引用
收藏
页码:270 / 278
页数:9
相关论文
共 9 条
  • [1] Bidez M W, 1992, Implant Dent, V1, P289, DOI 10.1097/00008505-199200140-00011
  • [2] BREEDING LC, 1993, INT J PROSTHODONT, V6, P435
  • [3] TIGHTENING CHARACTERISTICS FOR SCREWED JOINTS IN OSSEOINTEGRATED DENTAL IMPLANTS
    BURGUETE, RL
    JOHNS, RB
    KING, T
    PATTERSON, EA
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1994, 71 (06) : 592 - 599
  • [4] JEMT T, 1991, International Journal of Oral and Maxillofacial Implants, V6, P29
  • [5] A 3-YEAR FOLLOW-UP-STUDY ON SINGLE IMPLANT TREATMENT
    JEMT, T
    PETTERSSON, P
    [J]. JOURNAL OF DENTISTRY, 1993, 21 (04) : 203 - 208
  • [6] Jorneus L, 1992, Int J Oral Maxillofac Implants, V7, P353
  • [7] THE EFFECT OF CARBON GRAPHITE FIBER REINFORCEMENT ON THE STRENGTH OF PROVISIONAL CROWN AND FIXED PARTIAL DENTURE RESINS
    LARSON, WR
    DIXON, DL
    AQUILINO, SA
    CLANCY, JMS
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1991, 66 (06) : 816 - 820
  • [8] Shigley JE, 1977, MECH ENG DESIGN, V3, P240
  • [9] Shigley JE, 1986, STANDARD HDB MACHINE, P23