DOPPLER-ECHOCARDIOGRAPHIC RECOGNITION OF NORMAL REGURGITATION IN MECHANICAL HEART-VALVE PROSTHESES

被引:0
作者
FEHSKE, W [1 ]
KOHLER, J [1 ]
KESSEL, D [1 ]
RABAHIEH, R [1 ]
HAGENDORFF, A [1 ]
STEVENS, K [1 ]
NIEDEGGEN, A [1 ]
LUDERITZ, B [1 ]
机构
[1] RHEIN WESTFAL TH AACHEN,INST AERODYNAM,MED STROMUNGSTECHN LAB,W-5100 AACHEN,GERMANY
来源
ZEITSCHRIFT FUR KARDIOLOGIE | 1992年 / 81卷 / 11期
关键词
HEART VALVE PROSTHESES; LEAKAGE FLOW; DOPPLER-ECHOCARDIOGRAPHY; INVITRO TESTING; NORMAL VALUES;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The normal leakage flow in modem technical heart valve prostheses can be visualized by color-code Doppler-echocardiography, provided that an adequate ultra sonic image quality can be achieved. Sometimes it may be difficult, however, to distinguish such a normal leakage flow from pathological regurgitation. We therefore mounted new specimens of five different types of prostheses (Bjork-Shiley monostrut, Medtronic-Hall, Omnicarbon, Saint Jude Medical, Duromedics) into an invitro model, where the leakage flows could directly be visualized as emerging water jets. When the system was completely filled, the jets could also be registered by two-dimensional Doppler-echocardiography. For each valve, characteristic patterns for the localization of the principal jets were found. Besides the relative broad central jet in the Medtronic-Hall valve, all other jets arose mainly at the ring. They were detected at the hinges or the basis of the larger struts, respectively, and with asymmetrical mono-disc valves at the side of the smaller opening. The length and the orientation of the jets were found to change significantly with minimal variations of the position of the closing discs. This variability could be confirmed, when in a separate model the overall leakage flows were repeatedly measured. For patient examinations it seems useful to refer mainly to the localization of the jet origins. The comparison with reference values of jet-dimensions in Doppler images will rarely enable the examiner to distinguish between normal and pathological jets.
引用
收藏
页码:627 / 635
页数:9
相关论文
共 25 条
  • [1] COLOR FLOW DOPPLER EVALUATION OF ST-JUDE MEDICAL PROSTHETIC VALVES
    ALAM, M
    ROSMAN, HS
    MCBROOM, D
    GRAHAM, L
    MAGILLIGAN, DJ
    KHAJA, F
    STEIN, PD
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1989, 64 (19) : 1387 - 1389
  • [2] TRANSESOPHAGEAL COLOR FLOW DOPPLER AND ECHOCARDIOGRAPHIC FEATURES OF NORMAL AND REGURGITANT ST-JUDE-MEDICAL PROSTHESES IN THE MITRAL-VALVE POSITION
    ALAM, M
    SERWIN, JB
    ROSMAN, HS
    SHETH, M
    SUN, I
    SILVERMAN, NA
    GOLDSTEIN, S
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (10) : 871 - 873
  • [3] COLOR DOPPLER DIAGNOSIS OF MECHANICAL PROSTHETIC MITRAL REGURGITATION - USEFULNESS OF THE FLOW CONVERGENCE REGION PROXIMAL TO THE REGURGITANT ORIFICE
    BARGIGGIA, GS
    TRONCONI, L
    RAISARO, A
    RECUSANI, F
    RAGNI, T
    VALDESCRUZ, LM
    SAHN, DJ
    MONTEMARTINI, C
    [J]. AMERICAN HEART JOURNAL, 1990, 120 (05) : 1137 - 1142
  • [4] CHAMBERS J, 1989, BRIT HEART J, V62, P1
  • [5] DETECTION OF PROSTHETIC MITRAL-VALVE LEAK - A COMPARATIVE-STUDY USING TRANSESOPHAGEAL ECHOCARDIOGRAPHY, TRANSTHORACIC ECHOCARDIOGRAPHY, AND AUSCULTATION
    CHEN, YT
    KAN, MN
    CHEN, JS
    LIN, WW
    CHANG, MK
    HU, WS
    HWANG, DS
    LEE, DY
    HWANG, SL
    CHIANG, BN
    [J]. JOURNAL OF CLINICAL ULTRASOUND, 1990, 18 (07) : 557 - 561
  • [6] EVALUATION OF NORMAL PROSTHETIC VALVE FUNCTION BY DOPPLER ECHOCARDIOGRAPHY
    COOPER, DM
    STEWART, WJ
    SCHIAVONE, WA
    LOMBARDO, HP
    LYTLE, BW
    LOOP, FD
    SALCEDO, EE
    [J]. AMERICAN HEART JOURNAL, 1987, 114 (03) : 576 - 582
  • [7] DELLSPERGER KC, 1982, REGURGITATION PROSTH
  • [8] DIEBOLD B, 1983, BRIT HEART J, V49, P167
  • [9] TRANSESOPHAGEAL ECHOCARDIOGRAPHY IN THE EVALUATION OF PROSTHETIC AND NATIVE AORTIC VALVES
    DITTRICH, HC
    MCCANN, HA
    WALSH, TP
    BLANCHARD, DG
    OPPENHEIM, GE
    WAACK, TC
    DONAGHEY, LB
    WHEELER, K
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (07) : 758 - 761
  • [10] DITTRICH HC, 1988, AM HEART J, V1, P1