Characterization of CdMnTe radiation detectors using current and charge transients

被引:3
作者
Rafiei, R. [1 ,2 ]
Reinhard, M. I. [1 ]
Sarbutt, A. [1 ]
Uxa, S. [3 ]
Boardman, D. [1 ]
Watt, G. C. [1 ]
Belas, E. [3 ]
Kim, K. [4 ,5 ]
Bolotnikov, A. E. [4 ]
James, R. B. [4 ]
机构
[1] Australian Nucl Sci & Technol Org, Lucas Heights, NSW 2234, Australia
[2] Univ Western Australia, Sch Elect Elect & Comp Engn, Crawley, WA 6009, Australia
[3] Charles Univ Prague, Inst Phys, Fac Math & Phys, Ke Karlovu 5, CZ-12116 Prague 2, Czech Republic
[4] Brookhaven Natl Lab, Upton, NY 11973 USA
[5] Korea Univ, Dept Radiol Sci, Seoul 136703, South Korea
关键词
radiation detector; CdMnTe; CMT; detector fabrication; transient current technique; time of flight technique;
D O I
10.1088/1674-4926/34/7/073001
中图分类号
O469 [凝聚态物理学];
学科分类号
070205 ;
摘要
Charge transport characteristics of Cd0:95Mn0:05Te: In radiation detectors have been evaluated by combining time resolved current transient measurements with time of flight charge transient measurements. The shapes of the measured current pulses have been interpreted with respect to a concentration of net positive space-charge, which has resulted in an electric field gradient across the detector bulk. From the recorded current pulses the charge collection efficiency of the detector was found to approach 100%. From the evolution of the charge collection efficiency with applied bias, the electron mobility-lifetime product of mu n tau n =(8:5 +/- 0:4) 10(-4) cm(2)/V has been estimated. The electron transit time was determined using both transient current technique and time of flight measurements in the bias range of 100-1900 V. From the dependence of drift velocity on applied electric field the electron mobility was found to be mu n =(718 +/- 55) cm(2)/(V.s) at room temperature.
引用
收藏
页数:7
相关论文
共 37 条
[1]  
[Anonymous], 2010, MATLAB VERS 7 11 NA
[2]   Study of Te inclusions in CdMnTe crystals for nuclear detector applications [J].
Babalola, O. S. ;
Bolotnikov, A. E. ;
Groza, M. ;
Hossain, A. ;
Egarievwe, S. ;
James, R. B. ;
Burger, A. .
JOURNAL OF CRYSTAL GROWTH, 2009, 311 (14) :3702-3707
[3]   A comparison between irradiated magnetic Czochralski and float zone silicon detectors using the transient current technique [J].
Bates, AG ;
Moll, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 555 (1-2) :113-124
[4]   CHARGE-CARRIER MOBILITIES IN CD0.8ZN0.2TE SINGLE-CRYSTALS USED AS NUCLEAR RADIATION DETECTORS [J].
BURSHTEIN, Z ;
JAYATIRTHA, HN ;
BURGER, A ;
BUTLER, JF ;
APOTOVSKY, B ;
DOTY, FP .
APPLIED PHYSICS LETTERS, 1993, 63 (01) :102-104
[5]   ELECTRICAL-PROPERTIES AND PERFORMANCES OF NATURAL DIAMOND NUCLEAR RADIATION DETECTORS [J].
CANALI, C ;
GATTI, E ;
KOZLOV, SF ;
MANFREDI, PF ;
MANFREDOTTI, C ;
NAVA, F ;
QUIRINI, A .
NUCLEAR INSTRUMENTS & METHODS, 1979, 160 (01) :73-77
[6]   TRAP SPECTROSCOPY OF A-SI-H DIODES USING TRANSIENT CURRENT TECHNIQUES [J].
CRANDALL, RS .
JOURNAL OF ELECTRONIC MATERIALS, 1980, 9 (04) :713-726
[7]   Vertical Bridgman growth and characterization of CdMnTe crystals for gamma-ray radiation detector [J].
Du, Yuanyuan ;
Jie, Wanqi ;
Wang, Tao ;
Xu, Yadong ;
Yin, Liying ;
Yu, Pengfei ;
Zha, Gangqiang .
JOURNAL OF CRYSTAL GROWTH, 2011, 318 (01) :1062-1066
[8]   CARRIER DRIFT MOBILITY STUDY IN NEUTRON-IRRADIATED HIGH-PURITY SILICON [J].
EREMIN, V ;
LI, Z .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 362 (2-3) :338-343
[9]   Time of flight experimental studies of CdZnTe radiation detectors [J].
Erickson, JC ;
Yao, HW ;
James, RB ;
Hermon, H ;
Greaves, M .
JOURNAL OF ELECTRONIC MATERIALS, 2000, 29 (06) :699-703
[10]   Characterization of charge collection in CdTe and CZT using the transient current technique [J].
Fink, J. ;
Krueger, H. ;
Lodomez, P. ;
Wermes, N. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 560 (02) :435-443