IMPROVED TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT PATENCY WITH PTFE-COVERED STENT-GRAFTS - EXPERIMENTAL RESULTS IN SWINE

被引:114
|
作者
NISHIMINE, K
SAXON, RR
KICHIKAWA, K
MENDELHARTVIG, J
TIMMERMANS, HA
SHIM, HJ
UCHIDA, BT
BARTON, RE
KELLER, FS
ROSCH, J
机构
[1] OREGON HLTH SCI UNIV,SOTTER INTERVENT INST,PORTLAND,OR 97201
[2] NARA MED UNIV,DEPT RADIOL,NARA,JAPAN
[3] NARA MED UNIV,DEPT ONCORADIOL,NARA,JAPAN
[4] CHUNG ANG UNIV,DEPT RADIOL,SEOUL 156756,SOUTH KOREA
关键词
LIVER; INTERVENTIONAL PROCEDURE; SHUNTS; PORTOSYSTEMIC; STENTS AND PROSTHESES;
D O I
10.1148/radiology.196.2.7617843
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the ability of stent-grafts made with polytetrafluoroethylene (PTFE) graft material to improve transjugular intrahepatic portosystemic shunt (TIPS) patency. MATERIALS AND METHODS: TIPS were created in 13 swine by using PTFE-covered stent-grafts. Uncovered TIPS were placed in 13 other swine. Twenty-one of 26 animals were followed up with portal venography for 3 months or until the shunt became severely stenotic. Five animals without severe stenosis were sacrificed before 3 months because of illness. RESULTS: At 4 weeks after TIPS placement, nine of 13 stent-graft TIPS were patent (<50% diameter narrowing) compared with only one patent stent in 13 uncovered TIPS. Six of 13 stent-graft TIPS remained patent until the animals were sacrificed. In 11 of 12 uncovered TIPS, stenosis was most prominent in the parenchymal tract. In five of seven stent-graft TIPS, stenosis was most prominent in the hepatic vein above the end of the graft material. Bile leaks were discovered in six occluded uncovered TIPS and in two of the stent-graft TIPS. CONCLUSION: PTFE-covered stent-grafts significantly improved TIPS patency in swine (P < .01). However, stenosis in the hepatic vein led to late shunt malfunction in selected cases.
引用
收藏
页码:341 / 347
页数:7
相关论文
共 21 条
  • [1] Expanded PTFE-covered stent-grafts in the treatment of transjugular intrahepatic portosystemic shunt (TIPS) stenoses and occlusions
    Echenagusia, M
    Rodriguez-Rosales, G
    Simo, G
    Camuñez, F
    Bañares, R
    Echenagusia, A
    ABDOMINAL IMAGING, 2005, 30 (06): : 750 - 754
  • [2] Improved patency of transjugular intrahepatic portosystemic shunts in humans: Creation and revision with PTFE stent-grafts
    Haskal, ZJ
    RADIOLOGY, 1999, 213 (03) : 759 - 766
  • [3] Expanded PTFE-covered stent-grafts in the treatment of transjugular intrahepatic portosystemic shunt (TIPS) stenoses and occlusions
    M. Echenagusia
    G. Rodriguez-Rosales
    G. Simo
    F. Camuñez
    R. Bañares
    A. Echenagusia
    Abdominal Imaging, 2005, 30 : 750 - 754
  • [4] ePTFE-covered stent-grafts for revision of obstructed transjugular intrahepatic portosystemic shunt
    Cejna, M
    Peck-Radosavljevic, M
    Thurnher, S
    Schoder, M
    Rand, T
    Angermayr, B
    Lammer, J
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 25 (05) : 365 - 372
  • [5] Transjugular intrahepatic portosystemic shunt with silicone-covered wallstents: Results in a swine model
    Tanihata, H
    Saxon, RR
    Kubota, Y
    Pavcnik, D
    Uchida, BT
    Rosch, J
    Keller, FS
    Yamada, R
    Sato, M
    RADIOLOGY, 1997, 205 (01) : 181 - 184
  • [6] Transjugular intrahepatic portosystemic shunt with an autologous vein-covered stent: Results in a swine model
    Zhuang, ZW
    Hoopes, PJ
    Koutras, PC
    Ebbighausen, WH
    Wagner, RJ
    Bettmann, MA
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (11) : 1333 - 1342
  • [7] Primary implantation of polyester-covered stent-grafts for transjugular intrahepatic portosystemic stent shunts (TIPSS): A pilot study
    Cejna, M
    Thurnher, S
    Pidlich, J
    Kaserer, K
    Schoder, M
    Lammer, J
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 22 (04) : 305 - 310
  • [8] Creation of transjugular intrahepatic portosystemic shunts with stent-grafts: Initial experiences with a polytetrafluoroethylene-covered nitinol endoprosthesis
    Cejna, M
    Peck-Radosavljevic, M
    Thurnher, SA
    Hittmair, K
    Schoder, M
    Lammer, J
    RADIOLOGY, 2001, 221 (02) : 437 - 446
  • [9] Early and long-term clinical and radiological follow-up results of expanded-polytetrafluoroethylene-covered stent-grafts for transjugular intrahepatic portosystemic shunt procedures
    Geert Maleux
    Frederik Nevens
    Alexander Wilmer
    Sam Heye
    Chris Verslype
    Maria Thijs
    Guy Wilms
    European Radiology, 2004, 14 : 1842 - 1850
  • [10] Early and long-term clinical and radiological follow-up results of expanded-polytetrafluoroethylene-covered stent-grafts for transjugular intrahepatic portosystemic shunt procedures
    Maleux, G
    Nevens, F
    Wilmer, A
    Heye, S
    Verslype, C
    Thijs, M
    Wilms, G
    EUROPEAN RADIOLOGY, 2004, 14 (10) : 1842 - 1850