EFFECT OF CLOT-DETECTION METHODS AND REAGENTS ON ACTIVATED PARTIAL THROMBOPLASTIN TIME (APTT) - IMPLICATIONS IN HEPARIN MONITORING BY APTT

被引:60
作者
DANGELO, A
SEVESO, MP
DANGELO, SV
GILARDONI, F
DETTORI, AG
BONINI, P
机构
[1] OSPED SAN RAFFAELE,IST SCI,CENT ANAL LAB,I-20132 MILAN,ITALY
[2] OSPED REG PARMA,DIPARTIMENTO INTERDIV MED C CATTANI,CTR EMOSTASI & TROMBOSI,PARMA,ITALY
关键词
activated partial thromboplastin time; coagulometers; monitoring of heparin therapy;
D O I
10.1093/ajcp/94.3.297
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Two automatic coagulometers, ACL 810® (Instrumentation Laboratory), a laser-nephelometric centrifugal analyzer, and KoaguLab 40 A® (Ortho Diagnostics), an optical automatic coagulometer, were compared with the manual tilt-tube method for the performance of activated partial thromboplastin time (APTT). Seven commercial APTT reagents were used for duplicate determinations in 30 normal controls, 26 patients with liver disease, and 33 patients on full-dose heparin treatment. Clotting times were longer with the manual method than with ACL 810 and, to a lesser extent, with KoaguLab 40 A. Average imprecision of duplicate determinations (coefficient of variation [CV]) was less with ACL 810 (< 1.5%) than with KoaguLab 40 A (2.9%) and the manual method (2.4%). Differences in slope of the regression curves of clotting times obtained with the coagulometers over the tilt-tube method were observed with all the reagents tested (P < 0.01). Transformation of clotting times of controls, patients with liver disease, and patients on heparin therapy to APTT ratios did not eliminate the bias resulting from the different reagents (P < 0.001) and clot-detection methods (P < 0.001); in controls, significant (P < 0.001) reagent-method interaction was also observed. The in vitro heparin sensitivity differed with the APTT reagents evaluated and was influenced by the clot-detection method used. Transformation of APTT ratios of anticoagulated patients to apparent plasma heparin levels - as derived from in vitro dose-response curves - effectively eliminated the bias resulting from the different clot-detection methods but had no effect on the bias resulting from the different APTT reagents. In vitro heparin activity curves thus have little, if any, relevance for the ex vivo monitoring of heparin treatment.
引用
收藏
页码:297 / 306
页数:10
相关论文
共 34 条
  • [1] BAIN B, 1980, AM J CLIN PATHOL, V74, P668
  • [2] BANEZ EI, 1980, AM J CLIN PATHOL, V74, P569
  • [3] BECKALA HR, 1978, AM J CLIN PATHOL, V70, P71
  • [4] VARIABILITY IN HEPARIN SENSITIVITY OF APTT REAGENTS
    BJORNSSON, TD
    NASH, PV
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1986, 86 (02) : 199 - 204
  • [5] BRANDT JT, 1981, AM J CLIN PATHOL, V76, P530
  • [6] BRINKHOUSE K, 1980, CRC HDB SERIES CLIN, P227
  • [7] CHANTARANGKUL V, 1987, CLIN CHEM, V33, P1888
  • [8] CORNBLEET PJ, 1979, CLIN CHEM, V25, P432
  • [9] COMPARISON OF 2 AUTOMATED COAGULOMETERS AND THE MANUAL TILT-TUBE METHOD FOR THE DETERMINATION OF PROTHROMBIN TIME
    DANGELO, A
    SEVESO, MP
    DANGELO, SV
    GILARDONI, F
    MACAGNI, A
    MANOTTI, C
    BONINI, P
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 92 (03) : 321 - 328
  • [10] ANTICOAGULANTS IN VENOUS THROMBOEMBOLISM
    FENNERTY, A
    CAMPBELL, IA
    ROUTLEDGE, PA
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 1988, 297 (6659): : 1285 - 1288