IN-VITRO EVALUATION OF ENDOVASCULAR STENTS TO ASSESS SUITABILITY FOR ENDOVASCULAR GRAFT FIXATION

被引:20
作者
ANDREWS, SM [1 ]
ANSON, AW [1 ]
GREENHALGH, RM [1 ]
NOTT, DM [1 ]
机构
[1] CHARING CROSS HOSP,DEPT VASC SURG,LONDON W6 8RF,ENGLAND
关键词
ABDOMINAL AORTIC ANEURYSM; ENDOVASCULAR STENT; SHAPE MEMORY ALLOY;
D O I
10.1016/S1078-5884(05)80007-3
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: To compare and ascertain the suitability for endovascular aortic graft fixation of the Palmaz stent, the Strecker stent, the modified Gianturco-Z stent, the Wallstent and a new stent made of the shape memory alloy nitinol. Design, setting and materials: In vitro studies using a simple tensiometer to assess the forces required to distract each stent from porcine aorta; and using a high-pressure, pulsatile-flow pump to measure flow rates required to dislodge stents holding grafts within porcine aortas, with assessment using endovascular ultrasound Results: The Palmaz, the modified Gianturco-Z and the nitinol stents resisted significantly greater distraction forces than the Wallstent, but endovascular ultrasound examination revealed that the modified Gianturco-Z stent was seen to move away from the aortic wall at high flow rates precluding adequate fixation. The Strecker stent could not be assessed using these techniques. Conclusions: A nitinol stent may be suitable for aortic graft fixation, has characteristics similar to the Palmaz stent and has shape memory effect which may overcome the difficulties of introduction of wider diameter stents through narrow arteries.
引用
收藏
页码:403 / 407
页数:5
相关论文
共 5 条
  • [1] Parodi, Palmaz, Barone, Transfemoral intraluminal graft implantation for abdominal aortic aneurysm, Ann Vasc Surg, 5, pp. 491-499, (1991)
  • [2] Strecker, Liermann, Barth, Et al., Expandable tubular stents for treatment of arterial occlusive diseases: Experimental and clinical results, Radiology, 175, pp. 97-102, (1990)
  • [3] Chuter, Green, Ouriel, Fiore, DeWeese, Transfemoral endovascular aortic graft replacement, J Vasc Surg, 18, pp. 185-197, (1993)
  • [4] Zollikofer, Antonucci, Pfyffer, Et al., Arterial stent placement with use of the wallstent: Mid term results of clinical experience, Radiology, 179, pp. 449-456, (1991)
  • [5] Schetky, Shape-memory alloys, Sci Am, 241, pp. 74-83, (1979)