EXPERIMENTAL EVALUATION OF VELOCITY AND POWER ESTIMATION FOR ULTRASOUND BLOOD-FLOW IMAGING, BY MEANS OF A 2-DIMENSIONAL AUTOCORRELATION APPROACH

被引:143
作者
LOUPAS, T [1 ]
PETERSON, RB [1 ]
GILL, RW [1 ]
机构
[1] ADV TECHNOL LABS,BOTHELL,WA 98041
关键词
D O I
10.1109/58.393111
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper evaluates experimentally the performance of a novel axial velocity estimator, originally introduced in [4] and referred to as the 2D autocorrelator, and its Doppler power estimation counterpart, the 2D zero-lag autocorrelator, in the context of ultrasound color Row mapping, The evaluation also encompasses the well-established 1D autocorrelation technique for velocity estimation and its corresponding power estimator (1D zero-lag autocorrelator), to allow performance comparisons under identical conditions. Clutter-suppressed in vitro data sets from a steady-flow system are used to document the effect of the range gate and ensemble length, noise level and angle of insonation on the precision of the velocity estimates, The same data sets are used to examine issues related to the estimation of the Doppler signal's power, The first-order statistics of power estimates from regions corresponding to flow and noise are determined experimentally and the ability of power-based thresholding to separate flow signals from noise is characterized by means of ROC analysis, In summary, the results of the in vitro evaluation show that the proposed 2D-autocorrelatlon form of processing is consistently better than the corresponding 1D-autocorrelation techniques, in terms of both velocity and power estimation, Therefore, given their relatively modest implementation requirements, the 2D autocorrelation algorithms for velocity and power estimation appear to represent a superior, yet realistic, alternative to conventional Doppler processing for color flow mapping.
引用
收藏
页码:689 / 699
页数:11
相关论文
共 18 条