Submillimeter lung MRI at 0.55 T using balanced steady-state free precession with half-radial dual-echo readout (bSTAR)

被引:6
|
作者
Bauman, Grzegorz [1 ,2 ,5 ]
Lee, Nam G. G. [3 ]
Tian, Ye [4 ]
Bieri, Oliver [1 ,2 ]
Nayak, Krishna S. [3 ,4 ]
机构
[1] Univ Basel Hosp, Dept Radiol, Div Radiol Phys, Basel, Switzerland
[2] Univ Basel, Dept Biomed Engn, Basel, Switzerland
[3] Univ Southern Calif, Viterbi Sch Engn, Dept Biomed Engn, Los Angeles, CA USA
[4] Univ Southern Calif, Viterbi Sch Engn, Ming Hsieh Dept Elect & Comp Engn, Los Angeles, CA USA
[5] Univ Hosp Basel, Dept Radiol, Petersgraben 4, CH-4031 Basel, Switzerland
基金
美国国家科学基金会;
关键词
0; 55; T; balanced steady-state free precession; lung; non-Cartesian trajectories; radial MRI; TIME; PARENCHYMA;
D O I
10.1002/mrm.29757
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo demonstrate the feasibility of high-resolution morphologic lung MRI at 0.55 T using a free-breathing balanced steady-state free precession half-radial dual-echo imaging technique (bSTAR). MethodsSelf-gated free-breathing bSTAR (TE1/TE2/TR of 0.13/1.93/2.14 ms) lung imaging in five healthy volunteers and a patient with granulomatous lung disease was performed using a 0.55 T MR-scanner. A wobbling Archimedean spiral pole (WASP) trajectory was used to ensure a homogenous coverage of k-space over multiple breathing cycles. WASP uses short-duration interleaves randomly tilted by a small polar angle and rotated by a golden angle about the polar axis. Data were acquired continuously over 12:50 min. Respiratory-resolved images were reconstructed off-line using compressed sensing and retrospective self-gating. Reconstructions were performed with a nominal resolution of 0.9 mm and a reduced isotropic resolution of 1.75 mm corresponding to shorter simulated scan times of 8:34 and 4:17 min, respectively. Analysis of apparent SNR was performed in all volunteers and reconstruction settings. ResultsThe technique provided artifact-free morphologic lung images in all subjects. The short TR of bSTAR in conjunction with a field strength of 0.55 T resulted in a complete mitigation of off-resonance artifacts in the chest. Mean SNR values in healthy lung parenchyma for the 12:50 min scan were 3.6 +/- 0.8 and 24.9 +/- 6.2 for 0.9 mm and 1.75 mm reconstructions, respectively. ConclusionThis study demonstrates the feasibility of morphologic lung MRI with a submillimeter isotropic spatial resolution in human subjects with bSTAR at 0.55 T.
引用
收藏
页码:1949 / 1957
页数:9
相关论文
共 15 条
  • [1] Balanced steady-state free precession thoracic imaging with half-radial dual-echo readout on smoothly interleaved archimedean spirals
    Bauman, Grzegorz
    Bieri, Oliver
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (01) : 237 - 246
  • [2] Free-breathing half-radial dual-echo balanced steady-state free precession thoracic imaging with wobbling Archimedean spiral pole trajectories
    Bieri, Oliver
    Pusterla, Orso
    Bauman, Grzegorz
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (02): : 220 - 229
  • [3] Arterial spin labeled perfusion imaging with balanced steady-state free precession readout and radial sampling
    Han, Paul Kyu
    Marin, Thibault
    Zhuo, Yue
    Ouyang, Jinsong
    El Fakhri, Georges
    Ma, Chao
    MAGNETIC RESONANCE IMAGING, 2023, 102 : 126 - 132
  • [4] Cardiac balanced steady-state free precession MRI at 0.35 T: a comparison study with 1.5 T
    Rashid, Shams
    Han, Fei
    Gao, Yu
    Sung, Kyunghyun
    Cao, Minsong
    Yang, Yingli
    Hu, Peng
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2018, 8 (07) : 627 - 636
  • [5] Assessment of Small Bowel Motility Function With Cine-MRI Using Balanced Steady-State Free Precession Sequence
    Wakamiya, Makoto
    Furukawa, Akira
    Kanasaki, Shuzo
    Murata, Kiyoshi
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 33 (05) : 1235 - 1240
  • [6] Rapid Isotropic Resolution Cartilage Assessment Using Radial Alternating Repetition Time Balanced Steady-State Free-Precession Imaging
    Al Saleh, Habib
    Hernandez, Larry
    Lee, Kenneth S.
    Rosas, Humberto G.
    Block, Walter F.
    Kijowski, Richard
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 40 (04) : 796 - 803
  • [7] Feasibility of fetal cardiac function and anatomy assessment by real-time spiral balanced steady-state free precession magnetic resonance imaging at 0.55T
    Tian, Ye
    Detterich, Jon
    Pruetz, Jay D.
    Yagiz, Ecrin
    Wood, John C.
    Nayak, Krishna S.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2025, 27 (01)
  • [8] Rapid high-resolution T1 mapping using a highly accelerated radial steady-state free-precession technique
    Li, Zhitao
    Bilgin, Ali
    Johnson, Kevin
    Galons, Jean-Philippe
    Vedantham, Srinivasan
    Martin, Diego R.
    Altbach, Maria I.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2019, 49 (01) : 239 - 252
  • [9] Sequence comparison of spoiled gradient echo and balanced steady-state free precession for pulmonary free-breathing proton MRI in patients and healthy volunteers: Correspondence, repeatability, and validation with dynamic contrast-enhanced MRI
    Hahn, Jonah J.
    Voskrebenzev, Andreas
    Behrendt, Lea
    Klimes, Filip
    Poehler, Gesa H.
    Wacker, Frank
    Vogel-Claussen, Jens
    NMR IN BIOMEDICINE, 2024, 37 (11)
  • [10] Dynamic 2D and 3D mapping of hyperpolarized pyruvate to lactate conversion in vivo with efficient multi-echo balanced steady-state free precession at 3 T
    Mueller, Christoph A.
    Hundshammer, Christian
    Braeuer, Miriam
    Skinner, Jason G.
    Berner, Stephan
    Leupold, Jochen
    Duewel, Stephan
    Nekolla, Stephan G.
    Mansson, Sven
    Hansen, Adam E.
    von Elverfeldt, Dominik
    Ardenkjaer-Larsen, Jan H.
    Schilling, Franz
    Schwaiger, Markus
    Hennig, Juergen
    Hoevener, Jan-Bernd
    NMR IN BIOMEDICINE, 2020, 33 (06)