A 3RD-GENERATION WATER BATH BASED BLACKBODY SOURCE

被引:67
作者
FOWLER, JB
机构
关键词
APERTURE; BLACKBODY; CAVITY; CONICAL; EMISSIVITY; RADIATION; RADIOMETRY; REFLECTANCE; SOURCE; TEMPERATURE; THERMOMETER; WATER BATH;
D O I
10.6028/jres.100.044
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A third generation water bath based blackbody source has been designed and constructed in the Radiometric Physics Division at the National Institute of Standards and Technology, Gaithrrsburg, MD. The goal of this work was to design a large aperture blackbody source with improved temporal stability and reproducibility compared with earlier designs, as well as improved ease of use. These blackbody sources operate in the 278 K to 353 K range with water temperature combined standard uncertainties of 3.5 mK to 7.8 mK. The calculated emissivity of these sources is 0.9997 with a relative standard uncertainty of 0.0003. With a 50 mm limiting aperture at the cavity; entrance, the emissivity increases to 0.99997.
引用
收藏
页码:591 / 599
页数:9
相关论文
共 5 条
[1]  
GEIST J, 1971, APPL OPTICS, V10, P2188
[2]   THEORETICAL ANALYSIS OF LABORATORY BLACK-BODIES .1. GENERALIZED INTEGRAL-EQUATION [J].
GEIST, J .
APPLIED OPTICS, 1973, 12 (06) :1325-1330
[3]  
GEIST J, 1986, NBS1228 TECHN NOT
[4]   CALCULATION OF THE EFFECTIVE EMISSIVITIES OF SPECULAR-DIFFUSE CAVITIES BY THE MONTE-CARLO METHOD [J].
SAPRITSKY, VI ;
PROKHOROV, AV .
METROLOGIA, 1992, 29 (01) :9-14
[5]  
1994, MATERIALS SCI ENG HD