Characterization and validation of the thorax phantom Lungman for dose assessment in chest radiography optimization studies

被引:13
作者
Perez, Sunay Rodriguez [1 ,2 ]
Marshall, Nicholas William [3 ]
Struelens, Lara [1 ]
Bosmans, Hilde [3 ]
机构
[1] SCK CEN, Radiat Protect Dosimetry & Calibrat, Mol, Belgium
[2] Katholieke Univ Leuven, Med Phys & Qual Assessment, Leuven, Belgium
[3] Univ Hosp Gasthuisberg, Dept Radiol, Leuven, Belgium
关键词
anthropomorphic phantom; tissue equivalence; Monte Carlo simulations; optimization; digital radiology; chest radiography; organ dose;
D O I
10.1117/1.JMI.5.1.013504
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work concerns the validation of the Kyoto-Kagaku thorax anthropomorphic phantom Lungman for use in chest radiography optimization. The equivalence in terms of polymethyl methacrylate (PMMA) was established for the lung and mediastinum regions of the phantom. Patient chest examination data acquired under automatic exposure control were collated over a 2-year period for a standard x-ray room. Parameters surveyed included exposure index, air kerma area product, and exposure time, which were compared with Lungman values. Finally, a voxel model was developed by segmenting computed tomography images of the phantom and implemented in PENELOPE/penEasy Monte Carlo code to compare phantom tissue-equivalent materials with materials from ICRP Publication 89 in terms of organ dose. PMMA equivalence varied depending on tube voltage, from 9.5 to 10.0 cm and from 13.5 to 13.7 cm, for the lungs and mediastinum regions, respectively. For the survey, close agreement was found between the phantom and the patients' median values (deviations lay between 8% and 14%). Differences in lung doses, an important organ for optimization in chest radiography, were below 13% when comparing the use of phantom tissue-equivalent materials versus ICRP materials. The study confirms the value of the Lungman for chest optimization studies. (c) 2018 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:9
相关论文
共 36 条
  • [1] Method of simulating dose reduction for digital radiographic systems
    Båth, M
    Håkansson, M
    Tingberg, A
    Månsson, LG
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 253 - 259
  • [2] Low-voltage digital selenium radiography: Detection of simulated interstitial lung disease nodules and catheters - A phantom study
    Bernhardt, TM
    Rapp-Bernhardt, U
    Lenzen, H
    Roehl, FW
    Diederich, S
    Papke, K
    Ludwig, K
    Heindel, W
    [J]. RADIOLOGY, 2004, 232 (03) : 693 - 700
  • [3] Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV
    Boone, JM
    Seibert, JA
    [J]. MEDICAL PHYSICS, 1997, 24 (11) : 1661 - 1670
  • [4] DIGITAL CHEST RADIOGRAPHY WITH PHOTOSTIMULABLE STORAGE PHOSPHORS - SIGNAL-TO-NOISE RATIO AS A FUNCTION OF KILOVOLTAGE WITH MATCHED EXPOSURE RISK
    CHOTAS, HG
    FLOYD, CE
    DOBBINS, JT
    RAVIN, CE
    [J]. RADIOLOGY, 1993, 186 (02) : 395 - 398
  • [5] Design and application of a structured phantom for detection performance comparison between breast tomosynthesis and digital mammography
    Cockmartin, L.
    Marshall, N. W.
    Zhang, G.
    Lemmens, K.
    Shaheen, E.
    Van Ongeval, C.
    Fredenberg, E.
    Dance, D. R.
    Salvagnini, E.
    Michielsen, K.
    Bosmans, H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (03) : 758 - 780
  • [6] BEAM QUALITY INDEPENDENT ATTENUATION PHANTOM FOR ESTIMATING PATIENT EXPOSURE FROM X-RAY AUTOMATIC EXPOSURE CONTROLLED CHEST EXAMINATIONS
    CONWAY, BJ
    BUTLER, PF
    DUFF, JE
    FEWELL, TR
    GROSS, RE
    JENNINGS, RJ
    KOUSTENIS, GH
    MCCROHAN, JL
    RUETER, FG
    SHOWALTER, CK
    [J]. MEDICAL PHYSICS, 1984, 11 (06) : 827 - 832
  • [7] THRESHOLD PERCEPTION PERFORMANCE WITH COMPUTED AND SCREEN-FILM RADIOGRAPHY - IMPLICATIONS FOR CHEST RADIOGRAPHY
    DOBBINS, JT
    RICE, JJ
    BEAM, CA
    RAVIN, CE
    [J]. RADIOLOGY, 1992, 183 (01) : 179 - 187
  • [8] Chest radiography: Optimization of X-ray spectrum for cesium iodide-amorphous silicon flat-panel detector
    Dobbins, JT
    Samei, E
    Chotas, HG
    Warp, RJ
    Baydush, AH
    Floyd, CE
    Ravin, CE
    [J]. RADIOLOGY, 2003, 226 (01) : 221 - 230
  • [9] Dose-image quality optimisation in digital chest radiography
    Doyle, P
    Martin, CJ
    Gentle, D
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 269 - 272
  • [10] 3D Slicer as an image computing platform for the Quantitative Imaging Network
    Fedorov, Andriy
    Beichel, Reinhard
    Kalpathy-Cramer, Jayashree
    Finet, Julien
    Fillion-Robin, Jean-Christophe
    Pujol, Sonia
    Bauer, Christian
    Jennings, Dominique
    Fennessy, Fiona
    Sonka, Milan
    Buatti, John
    Aylward, Stephen
    Miller, James V.
    Pieper, Steve
    Kikinis, Ron