VARIABLES AFFECTING THE INTRODUCTION OF LARGE-SAMPLE VOLUMES IN CAPILLARY GAS-CHROMATOGRAPHY USING A PROGRAMMED-TEMPERATURE VAPORIZER

被引:36
作者
SENORANS, FJ
TABERA, J
VILLEN, J
HERRAIZ, M
REGLERO, G
机构
[1] CSIC,INST FERMENTAC IND,JUAN CIERVA 3,E-28006 MADRID,SPAIN
[2] UNIV CASTILLA LA MANCHA,ESCUELA POLITECN ALBACETE,E-02006 ALBACETE,SPAIN
来源
JOURNAL OF CHROMATOGRAPHY | 1993年 / 648卷 / 02期
关键词
D O I
10.1016/0021-9673(93)80423-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The dependence of the programmed-temperature vaporizer (PTV) solvent split sampling technique on eight experimental variables affecting the introduction of large sample volumes in capillary GC was evaluated. The simplex method was used to improve the sensitivity of the analysis significantly by optimization of the operating conditions, namely sample volume, flow-rate during sampling, initial PTV temperature, length of the adsorbent bed, nature of the trapping material used in the glass liner, speed of sample introduction, elimination volume and solvent elimination temperature. With this method, reliable determinations of several volatile components at sub-ppb concentrations were obtained using a flame ionization detector, without the need for prior enrichment of the sample.
引用
收藏
页码:407 / 414
页数:8
相关论文
共 35 条
[1]   AUTOMATED OPTIMIZATION IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
BERRIDGE, JC .
ANALYTICA CHIMICA ACTA, 1986, 191 :243-259
[2]   PRECONCENTRATION OF VOLATILE COMPONENTS OF FOODS - OPTIMIZATION OF THE STEAM DISTILLATION SOLVENT-EXTRACTION AT NORMAL PRESSURE [J].
BLANCH, GP ;
TABERA, J ;
HERRAIZ, M ;
REGLERO, G .
JOURNAL OF CHROMATOGRAPHY, 1993, 628 (02) :261-268
[3]   DIRECT SAMPLE INTRODUCTION OF HIGH BOILING COMPOUNDS ONTO GLASS CAPILLARY COLUMNS - COMPARISON OF MANUAL AND AUTOMATIC SAMPLING [J].
CRAMERS, CA ;
VERMEER, EA .
CHROMATOGRAPHIA, 1975, 8 (09) :479-481
[4]   SOLVENT-FREE AND SPLITLESS INJECTION METHOD FOR OPEN TUBULAR COLUMNS [J].
DELEEUW, JW ;
MATERS, WL ;
MEENT, DVD ;
BOON, JJ .
ANALYTICAL CHEMISTRY, 1977, 49 (12) :1881-1884
[5]  
DEMING SN, 1973, ANAL CHEM, V45, pA278
[6]  
FISHER RJ, 1989, FOOD TECHNOL-CHICAGO, V43, P90
[7]  
GROB K, 1974, J CHROMATOGR, V94, P53
[8]   SPLITLESS INJECTION ON CAPILLARY COLUMNS, .I. BASIC TECHNIUE - STEROID ANALYSIS AS AN EXAMPLE [J].
GROB, K ;
GROB, G .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1969, 7 (10) :584-&
[9]  
GROB K, 1990, HRC-J HIGH RES CHROM, V13, P540
[10]   SPLITLESS INJECTION ON CAPILLARY COLUMNS, .2. CONDITIONS AND LIMITS, PRACTICAL REALIZATION [J].
GROB, K ;
GROB, G .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1969, 7 (10) :587-&