Accuracy of implant surgery with surgical guide by inexperienced clinicians: an in vitro study

被引:13
作者
Toyoshima, Takeshi [1 ]
Tanaka, Hideaki [1 ,2 ]
Sasaki, Masanori [2 ]
Ichimaru, Eiji [3 ]
Naito, Yasushi [3 ]
Matsushita, Yasuyuki [2 ,4 ]
Koyano, Kiyoshi [2 ,4 ]
Nakamura, Seiji [1 ]
机构
[1] Kyushu Univ, Fac Dent Sci, Div Maxillofacial Diagnost & Surg Sci, Fukuoka, Japan
[2] Kyushu Univ Hosp, Regenerat Dent & Implant Ctr, Fukuoka, Japan
[3] Shin Tosu Periodontal & Implant Dent Ctr, Ichimaru Dental Off Kuranoue, Tosu, Saga, Japan
[4] Kyushu Univ, Fac Dent Sci, Div Oral Rehabil, Fukuoka, Japan
关键词
artificial mandible; inexperienced clinician; in vitro; implant surgery; model; surgical guide; unilateral free-end edentulism;
D O I
10.1002/cre2.3
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Implant surgery with surgical guide has been introduced with a concept of position improvement. The surgery might be considered as easy even for inexperienced clinician because of step simplicity. However, there are residual risks, resulting in postoperative complications. The aim of this study was to assess the accuracy of implant surgery with surgical guide by inexperienced clinicians in in vitro. After preoperative computed tomographies (CTs) of five artificial models of unilateral free-end edentulism with scan templates, five surgical guides were established from templates. Following virtual planning, 10 implants were placed in the 45 and 47 regions by five residents without any placement experiences. All drillings and placements were performed using surgical guides. After postoperative CTs, inaccurate verifications between virtual and actual positions of implants were carried out, by overlaying of pre/postoperative CT data. The angle displacement of implant axis in the 47 region was significantly larger than that in the 45 region (P= 0.031). The 3D offset of implant base in the 47 region was significantly larger than that in the 45 region (P= 0.002). For distal/apical directions, displacements of base in the 47 region were significantly larger than those in the 45 region (P= 0.004 and P= 0.003, respectively). The 3D offset of implant tip in the 47 region was significantly larger than that in the 45 region (P= 0.003). For distal/apical directions, displacements of tip in the 47 region were significantly larger than those in the 45 region (P= 0.002 and P= 0.003, respectively). Within limitations of this in vitro study, data for accuracy of implant surgery with surgical guide would be informative for further studies, because in vitro studies should be substantially made to avoid unnecessary burden of patients, in advance of retro/prospective studies. A comparison of the accuracy in this in vitro model between by inexperienced and well-experienced operators should be necessary for clinicians intending to use surgical guide for placement.
引用
收藏
页码:10 / 17
页数:8
相关论文
共 26 条
[1]   Accuracy of Two Stereolithographic Guide Systems for Computer-Aided Implant Placement: A Computed Tomography-Based Clinical Comparative Study [J].
Arisan, Volkan ;
Karabuda, Z. Cuneyt ;
Ozdemir, Tayfun .
JOURNAL OF PERIODONTOLOGY, 2010, 81 (01) :43-51
[2]   Root Contact during Drilling for Microimplant Placement Affect of Surgery Site and Operator Expertise [J].
Cho, U-Hyeong ;
Yu, Wonjae ;
Kyung, Hee-Moon .
ANGLE ORTHODONTIST, 2010, 80 (01) :130-136
[3]   IMPACT OF OPERATOR EXPERIENCE ON THE ACCURACY OF IMPLANT PLACEMENT WITH STEREOLITHOGRAPHIC SURGICAL TEMPLATES: AN IN VITRO STUDY [J].
Cushen, Sarra E. ;
Turkyilmaz, Ilser .
JOURNAL OF PROSTHETIC DENTISTRY, 2013, 109 (04) :248-254
[4]   A model study on flapless implant placement by clinicians with a different experience level in implant surgery [J].
de Velde, Tommie Van ;
Glor, Fadi ;
De Bruyn, Hugo .
CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (01) :66-72
[5]   Basic research and 12 years of clinical experience in computer-assisted navigation technology: a review [J].
Ewers, R ;
Schicho, K ;
Undt, G ;
Wanschitz, F ;
Truppe, M ;
Seemann, R ;
Wagner, A .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2005, 34 (01) :1-8
[6]  
Hammerle CHF, 2009, INT J ORAL MAX IMPL, V24, P126
[7]  
Jung RE, 2009, INT J ORAL MAX IMPL, V24, P92
[8]  
Kohavi David, 2004, J Oral Implantol, V30, P23, DOI 10.1563/1548-1336(2004)030<0023:RCRODI>2.0.CO
[9]  
2
[10]   Virtually planned and template-guided implant surgery: an experimental model matching approach [J].
Komiyama, Ai ;
Pettersson, Andreas ;
Hultin, Margareta ;
Nasstrom, Karin ;
Klinge, Bjorn .
CLINICAL ORAL IMPLANTS RESEARCH, 2011, 22 (03) :308-313