AN IN-VITRO EVALUATION OF VENOUS CANNULA IN A SIMULATED PARTIAL (FEMOROFEMORAL) CARDIOPULMONARY BYPASS CIRCUIT

被引:6
作者
KUGAI, T
KOJA, K
KUSABA, A
机构
[1] Second Department of Surgery, Faculty of Medicine, University of the Ryukyus, Okinawa
关键词
PARTIAL CARDIOPULMONARY BYPASS; OPTIMAL BYPASS FLOW; MODEL CIRCUIT;
D O I
10.1111/j.1525-1594.1995.tb02305.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We designed a study to evaluate three factors (siphon gradient [PH], the right atrial pressure [RAP], and the inferior vena caval flow [IVCF]) to be optimized to maximize the venous drainage flow (DF) during partial cardiopulmonary bypass using eight venous cannulas of three different types and an original model circuit. The relationship between venous DF and the three factors is indicated by the multiple regression equation DF2 = alpha PH + beta RAP + gamma IVCF2 + C, where alpha, beta, and gamma are regression estimates and C is a constant. Multiple regression analysis results showed that DF was positively correlated with PH and RAP and negatively correlated with IVCF. A long cannula with 12 side holes and 60 cm long was considered to be useful to yield the optimal venous drainage flow under the condition of maintenance of the flow balance (DF and ICVF) and the pressure balance (RAP and IVCP) at the zero point. Moreover, this model may allow extensive research in flow dynamics of venous cannula without involving human subjects.
引用
收藏
页码:154 / 160
页数:7
相关论文
共 22 条
[1]   A NEW VENOUS CANNULA FOR USE DURING CPB [J].
BAHADIR, I ;
DIETHRICH, EB .
ANNALS OF THORACIC SURGERY, 1984, 37 (05) :439-440
[2]  
CARTIER R, 1990, J THORAC CARDIOV SUR, V99, P1038
[3]  
Dillon M L, 1967, Ann Thorac Surg, V3, P430
[4]  
HIRAYAMA T, 1991, Journal of the Japanese Association for Thoracic Surgery, V39, P1153
[5]  
Kawashima Y., 1992, Japanese Journal of Artificial Organs, V21, P1457
[6]   SURGICAL-TREATMENT OF ANEURYSMS OF THE THORACIC AORTA WITH THE AID OF PARTIAL CARDIOPULMONARY BYPASS - AN ANALYSIS OF 95 PATIENTS [J].
KAZUI, T ;
KOMATSU, S ;
YOKOYAMA, H .
ANNALS OF THORACIC SURGERY, 1987, 43 (06) :622-627
[7]  
KAZUI T, 1980, JPN THORAC SURG, V28, P1752
[8]  
KOLOBOW T, 1970, SURGERY, V68, P625
[9]   FLOW THROUGH COLLAPSIBLE TUBES AT LOW REYNOLDS-NUMBERS - APPLICABILITY OF THE WATERFALL MODEL [J].
LYON, CK ;
SCOTT, JB ;
WANG, CY .
CIRCULATION RESEARCH, 1980, 47 (01) :68-73
[10]  
MONTOYA JP, 1991, T ASAIO, V37, P4