OPTIMIZING TECHNICAL CONDITIONS FOR MAGNETIC-RESONANCE-IMAGING OF THE RAT-BRAIN AND ABDOMEN IN A LOW MAGNETIC-FIELD

被引:9
|
作者
YAMADA, K
MIYAHARA, K
SATO, M
HIROSE, T
YASUGI, Y
MATSUDA, Y
FURUHAMA, K
机构
[1] HITACHI MED CO,DIV ENGN,CHIBA 277,JAPAN
[2] DAIICHI PHARMACEUT CO LTD,EXPLORATORY RES LABS 3,TOKYO 136,JAPAN
关键词
MRI; RAT; LOW MAGNETIC FIELD;
D O I
10.1111/j.1740-8261.1995.tb00306.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The present study was designed to establish the appropriate technical conditions for magnetic resonance imaging (MRI) of the head and abdomen in rats using a low magnetic field strength (0.2 T) MRI unit equipped with three radio frequency (RF) coils: a custom-made solenoid coil, a temporomandibular joint surface coil and a knee quadrature (QD) RF coil, Male adult Sprague-Dawley rats were used, T1 and T2 relaxation times of both anatomic regions were measured from T1 and T2 maps, An appropriate parameter was then used to make final T1 and T2 weighted images, It was found that the most suitable coil for the head was the custom-made solenoid coil, and that for the abdomen, the knee QD coil, The T1 and T2 relaxation times were 314 to 316 msec and 72 to 74 msec for the head, and 220 to 252 msec and 42 to 51 msec for the abdomen. The optimum parameters for the head were TR/TE = 400/38 msec in T1 weighted images and TR/TE = 1,800/110 msec in T2 weighted images, and for the abdomen, TRITE = 300/25 msec in T1 weighted images and TR/TE = 1,500/110 msec in T2 weighted images, These results demonstrate that a low magnetic field strength MRI unit has potential for MRI study of the brain and abdomen in rats.