Stability of a new, multichannel, low-level liquid scintillation counter system, Kvartett

被引:3
作者
Einarsson, S [1 ]
Theodorsson, P [1 ]
机构
[1] UNIV ICELAND,INST SCI,IS-107 REYKJAVIK,ICELAND
关键词
D O I
10.1017/S003382220003126X
中图分类号
P3 [地球物理学]; P59 [地球化学];
学科分类号
0708 ; 070902 ;
摘要
Kvartett is a new liquid scintillation counting (LSC) system for radiocarbon dating that takes a radical departure from conventional systems to obtain a compact, low-level counting system measuring four samples simultaneously. Each sample vial, inside the well of a large NaI(Tl) guard-counter crystal (facing down), sits on top of a vertical PMT, The fourfold counting capacity can be used to increase the number of samples being dated or to get higher precision. The increased throughput helps to keep a rigid quality-control standard. We monitored the background count rate almost continuously for 7 months, and measured the count rate of a standard repeatedly for 2 months. The results show the background and system reproducibility to be stable.