CLINICAL-EVALUATION OF AN AUTOMATIC SENSITIVITY ADJUSTMENT FEATURE IN A DUAL CHAMBER PACEMAKER

被引:15
|
作者
WILSON, JH
LOVE, CJ
WETTENSTEIN, EH
机构
[1] OHIO STATE UNIV,COLUMBUS,OH 43210
[2] INTERMED INC,FREEPORT,TX
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 10期
关键词
automatic sensitivity adjustment; dual chamber pacemaker;
D O I
10.1111/j.1540-8159.1990.tb02017.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The clinical evaluation of a new automatic sensitivity adjustment feature in the Cosmos fl, Model #284–05 is described. This feature, designed to maintain a 2:1 safety margin for sensing intrinsic signals, was activated by way of special programmer software in ten patients at two centers. While the feature worked satisfactorily in some cases, it did not perform as expected, and undersensing in both chambers was observed. This may have been due to biasing the adjustment toward maximum rather than minimum electrograms, This study suggested that electrograms may vary by more than 100% which underscores the importance of this feature, and reinforces the need for continued development. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:1220 / 1223
页数:4
相关论文
共 6 条
  • [1] Automatic atrial anti-tachy pacing for the termination of spontaneous atrial tachyarrhythmias: Clinical experience with a novel dual-chamber pacemaker
    Vollmann, D
    Stevens, J
    Buchwald, AB
    Unterberg, C
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2001, 5 (04) : 477 - 485
  • [2] Automatic Atrial Anti-Tachy Pacing for the Termination of Spontaneous Atrial Tachyarrhythmias: Clinical Experience with a Novel Dual-Chamber Pacemaker
    Dirk Vollmann
    Jens Stevens
    Arnd B. Buchwald
    Christina Unterberg
    Journal of Interventional Cardiac Electrophysiology, 2001, 5 : 477 - 485
  • [3] Evaluation of left ventricular longitudinal function and synchrony after dual chamber pacemaker implantation
    Baronaite-Dadoniene, Kristina
    Vaskelyte, Jolanta
    Puodziukynas, Aras
    Zabiela, Vytautas
    Kazakevicius, Tomas
    Sakalyte, Gintare
    MEDICINA-LITHUANIA, 2014, 50 (06): : 340 - 344
  • [4] Clinical implications of elective replacement indicator setting changes in patients with dual-chamber pacemaker devices
    Phillips, Kari A.
    Ponamgi, Shiva
    Mundell, Benjamin
    Krushelnytskyy, Mykhaylo
    Li, Zhuo
    Rea, Robert
    Deshmukh, Abhishek
    McLeod, Christopher
    Espinosa, Raul E.
    Osborn, Michael
    Friedman, Paul A.
    Mulpuru, Siva K.
    Cha, Yong-Mei
    Neutzling, Lori B.
    Munger, Thomas
    Kancharla, Krishna
    Asirvatham, Samuel J.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (10) : 2704 - 2710
  • [5] Automatic measurement of atrial pacing thresholds in dual-chamber pacemakers: Clinical experience with atrial capture management
    Sperzel, J
    Milasinovic, G
    Smith, TW
    Mead, H
    Brandt, J
    Haisty, WK
    Bailey, JR
    Roelke, M
    Simonson, J
    Englund, J
    Farges, E
    Compton, S
    HEART RHYTHM, 2005, 2 (11) : 1203 - 1210
  • [6] Clinical benefit of atrio-ventricular delay optimization in patients with a dual-chamber pacemaker: a pilot study. The CBRAVO trial (NCT01998256)
    Cools, Thijs
    Daerden, Tara
    Herbots, Lieven
    Geukens, Robert
    Verwerft, Jan
    Koopman, Pieter
    Dilling-Boer, Dagmara
    Hansen, Dominique
    Vranckx, Pascal
    Dendale, Paul
    ACTA CARDIOLOGICA, 2016, 71 (03) : 257 - 265