Evaluation of an iterative model-based reconstruction algorithm for low-tube-voltage (80 kVp) computed tomography angiography

被引:21
作者
Noel, Peter B. [1 ]
Koehler, Thomas [2 ]
Fingerle, Alexander A. [1 ]
Brown, Kevin M. [3 ]
Zabic, Stanislav [3 ]
Muenzel, Daniela [1 ]
Haller, Bernhard [4 ]
Baum, Thomas [1 ]
Henninger, Martin [1 ]
Meier, Reinhard [1 ]
Rummeny, Ernst J. [1 ]
Dobritz, Martin [1 ]
机构
[1] Tech Univ Munich, Dept Radiol, D-81675 Munich, Germany
[2] Philips Res Labs, D-22335 Hamburg, Germany
[3] Philips Healthcare, Cleveland, OH 44143 USA
[4] Tech Univ Munich, Inst Med Stat & Epidemiol, D-81675 Munich, Germany
关键词
computed tomography; iterative reconstruction; low-tube-current; computed tomography angiography; dose reduction; low-tube-voltage;
D O I
10.1117/1.JMI.1.3.033501
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The objective of this study was to investigate the improvement in diagnostic quality of an iterative model-based reconstruction (IMBR) algorithm for low-tube-voltage (80-kVp) and low-tube-current in abdominal computed tomography angiography (CTA). A total of 11 patients were imaged on a 256-slice multidetector computed tomography for visualization of the aorta. For all patients, three different reconstructions from the low-tube-voltage data are generated: filtered backprojection (FBP), IMBR, and a mixture of both IMBR + FBP. To determine the diagnostic value of IMBR-based reconstructions, the image quality was assessed. With IMBRbased reconstructions, image noise could be significantly reduced, which was confirmed by a highly improved contrast-to-noise ratio. In the image quality assessment, radiologists were able to reliably detect more third-order and higher aortic branches in the IMBR reconstructions compared to FBP reconstructions. The effective dose level was, on average, 3.0 mSv for 80-kVp acquisitions. Low-tube-voltage CTAs significantly improve vascular contrast as presented by others; however, this effect in combination with IMBR enabled yet another substantial improvement of diagnostic quality. For IMBR, a significant improvement of image quality and a decreased radiation dose at low-tube-voltage can be reported. (C) The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License.
引用
收藏
页数:7
相关论文
共 24 条
[1]   Radiation Exposure and Image Quality of Normal Computed Tomography Brain Images Acquired With Automated and Organ-Based Tube Current Modulation Multiband Filtering and Iterative Reconstruction [J].
Becker, Hans-Christoph ;
Augart, Dominik ;
Karpitschka, Martina ;
Ulzheimer, Stefan ;
Bamberg, Fabian ;
Morhard, Dominik ;
Neumaier, Klement ;
Graser, Anno ;
Johnson, Thorsten ;
Reiser, Maximilian .
INVESTIGATIVE RADIOLOGY, 2012, 47 (03) :202-207
[2]  
Bergner F, 2012, P 2 INT C IM FORM XR, P267
[3]   Acceleration of ML Iterative Algorithms for CT by the use of Fast Start Images [J].
Brown, Kevin M. ;
Zabic, Stanislav ;
Koehler, Thomas .
MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
[4]  
Fessler J. A., 2000, STAT IMAGE RECONSTRU
[5]   From the RSNA refresher courses - CT angiography: Clinical applications in the abdomen [J].
Fishman, EK .
RADIOGRAPHICS, 2001, 21 :S3-S16
[6]   Impact of iterative reconstruction on CNR and SNR in dynamic myocardial perfusion imaging in an animal model [J].
Gramer, B. M. ;
Muenzel, D. ;
Leber, V. ;
von Thaden, A. -K. ;
Feussner, H. ;
Schneider, A. ;
Vembar, M. ;
Soni, N. ;
Rummeny, E. J. ;
Huber, A. M. .
EUROPEAN RADIOLOGY, 2012, 22 (12) :2654-2661
[7]   Assessment of an iterative reconstruction algorithm (SAFIRE) on image quality in pediatric cardiac CT datasets [J].
Han, B. Kelly ;
Grant, Katharine L. R. ;
Garberich, Ross ;
Sedlmair, Martin ;
Lindberg, Jana ;
Lesser, John R. .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2012, 6 (03) :200-204
[8]   CT angiography with volume rendering: Advantages and applications in splanchnic vascular imaging [J].
Johnson, PT ;
Heath, DG ;
Kuszyk, BS ;
Fishman, EK .
RADIOLOGY, 1996, 200 (02) :564-568
[9]   64-Slice CT angiography of the abdominal aorta and abdominal arteries: comparison of the diagnostic efficacy of iobitridol 350 mgI/ml versus iomeprol 400 mgI/ml in a prospective, randomised, double-blind multi-centre trial [J].
Loewe, Christian ;
Becker, Christoph R. ;
Berletti, Riccardo ;
Cametti, Carlo Alberto ;
Caudron, Jerome ;
Coudyzer, Walter ;
De Mey, Johan ;
Favat, Massimo ;
Heautot, Jean-Francois ;
Heye, Sam ;
Hittinger, Markus ;
Larralde, Antoine ;
Lestrat, Jean-Pierre ;
Marangoni, Roberto ;
Nieboer, Koenraad ;
Reimer, Peter ;
Schwarz, Martin ;
Schernthaner, Melanie ;
Lammer, Johannes .
EUROPEAN RADIOLOGY, 2010, 20 (03) :572-583
[10]   Low-Tube-Voltage, High-Tube-Current Multidetector Abdominal CT: Improved Image Quality and Decreased Radiation Dose with Adaptive Statistical Iterative Reconstruction Algorithm-Initial Clinical Experience [J].
Marin, Daniele ;
Nelson, Rendon C. ;
Schindera, Sebastian T. ;
Richard, Samuel ;
Youngblood, Richard S. ;
Yoshizumi, Terry T. ;
Samei, Ehsan .
RADIOLOGY, 2010, 254 (01) :145-153