A SEGMENTED TIME-OF-FLIGHT MASS-SPECTROMETER

被引:3
作者
SRIVASTAVA, SK
IGA, I
RAO, MVVS
机构
[1] Jet Propulsion Lab., California Inst. of Technol., Pasadena, CA
关键词
D O I
10.1088/0957-0233/6/9/020
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The present paper describes the design of a time-of-flight mass spectrometer (TOFMS) in which the single flight tube of a conventional TOFMS has been replaced by several cylindrical electrostatic lenses in tandem. By a judicious choice of voltages on these lenses, an improved TOFMS has been realized which has a superior mass and energy resolution, shorter flight lengths, excellent signal-to-noise ratio and less stringent requirements on the bias voltages.
引用
收藏
页码:1379 / 1382
页数:4
相关论文
共 4 条
[1]  
BELL RE, 1966, ALPHA BETA GAMMA RAY, V2, P911
[2]   CROSS-SECTIONS FOR THE PRODUCTION OF N-2+, N+ + N22+ AND N-2+ BY ELECTRON-IMPACT ON N-2 [J].
KRISHNAKUMAR, E ;
SRIVASTAVA, SK .
JOURNAL OF PHYSICS B-ATOMIC MOLECULAR AND OPTICAL PHYSICS, 1990, 23 (11) :1893-1903
[3]   THE RENAISSANCE OF TIME-OF-FLIGHT MASS-SPECTROMETRY [J].
PRICE, D ;
MILNES, GJ .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1990, 99 (1-2) :1-39
[4]   TIME-OF-FLIGHT MASS SPECTROMETER WITH IMPROVED RESOLUTION [J].
WILEY, WC ;
MCLAREN, IH .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1955, 26 (12) :1150-1157