IN-VITRO AND IN-VIVO VALIDATION TESTS FOR TOTAL ARTIFICIAL-HEART

被引:17
作者
ORIME, Y
TAKATANI, S
TASAI, K
OHARA, Y
NAITO, K
MIZUGUCHI, K
MAKINOUCHI, K
MATSUDA, Y
SHIMONO, T
GLUECK, J
NOON, GP
NOSE, Y
机构
[1] Department of Surgery, Baylor College of Medicine, Houston, Texas
关键词
TOTAL ARTIFICIAL HEART; TOTALLY IMPLANTABLE ARTIFICIAL HEART; IN-VITRO TEST; IN-VIVO TEST;
D O I
10.1111/j.1525-1594.1994.tb03299.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Properly planned in vitro and in vivo studies are mandatory to validate blood pump performance. However, standard procedures for in vitro and in vivo tests of total artificial heart (TAH) performance still do not exist. At Baylor College of Medicine, standard tests procedures for performance validation of the totally implantable TAH have been established. In this paper, these in vitro tests protocols (performance mapping tests, accelerated endurance test, hemolysis test, flow visualization tests, etc.) are described as well as in vivo test procedures (TAH implantation procedure, including surgical technique, postoperative management and tests, etc.). In addition, pathological protocols are presented. In our experience, these procedures are very simple, easy to perform, and inexpensive. These protocols are proposed as standard in vitro and in vivo tests for a TAH.
引用
收藏
页码:54 / 72
页数:19
相关论文
共 24 条
[1]  
EMOTO H, 1991, ARTIF ORGANS, V15, P29
[2]  
HARASAKI H, 1986, 1986 DEV TECHN BRANC
[3]  
HARASAKI H, 1985, T SOC BIOMAT, V8, P78
[4]  
LEVINSON M M, 1986, Journal of Heart Transplantation, V5, P236
[5]  
MOISE J, 1982, DEV IMPLANTABLE ELEC
[6]  
NOSE Y, 1992, ARTIF ORGANS, V16, P19
[7]  
NOSE Y, 1975, NHLBI1HV42960 REP
[8]  
NOSE Y, 1979, NHLBI1HV429605 REP, P75
[9]  
ORIME Y, 1992, ARTIF ORGANS, V16, P607
[10]  
ORIME Y, 1993, T AM SOC ART INT ORG, V39, pM172