A COMPARISON OF TRANSCUTANEOUS, END-TIDAL AND ARTERIAL MEASUREMENTS OF CARBON-DIOXIDE DURING GENERAL-ANESTHESIA

被引:44
|
作者
REID, CW
MARTINEAU, RJ
MILLER, DR
HULL, KA
BAINES, J
SULLIVAN, PJ
机构
[1] OTTAWA GEN HOSP, DEPT ANAESTHESIA, 501 SMYTH RD, OTTAWA K1H 8L6, ONTARIO, CANADA
[2] OTTAWA GEN HOSP, DEPT BIOMED ENGN, OTTAWA K1H 8L6, ONTARIO, CANADA
[3] UNIV OTTAWA, OTTAWA K1N 6N5, ONTARIO, CANADA
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1992年 / 39卷 / 01期
关键词
CARBON DIOXIDE; MONITORING; TENSION; ARTERIAL; CUTANEOUS; END-TIDAL; GRADIENTS; VENTILATION; CARBON DIOXIDE TENSION;
D O I
10.1007/BF03008669
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A randomized, prospective study was performed to evaluate the accuracy of a new transcutaneous carbon dioxide (CO2) monitor (Fastrac(TM)) during general anaesthesia. Twenty-two adult patients undergoing elective surgery were subjected to three different levels of minute ventilation by varying their respiratory rates in a randomized cross-over design. Simultaneous measurements of transcutaneous CO2 (PTCCO2) and arterial CO2 (PaCO2) were obtained at three levels of minute ventilation (low, medium and high). End-tidal CO2 (PETCO2) values were also recorded from a mass spectrometer (SARA(TM)) at each time period. A total of 66 data sets with PaCO2 ranging from 28-62 mmHg were analyzed. The PTCCO2 values demonstrated a high degree of correlation with PaCO2 over the range of minute ventilation (y = 0.904x + 6.36, r = 0.92, P < 0.001). The PETCO2 measurement also demonstrated a generally good correlation with PaCO2 (y = 0.62x + 9.21, r = 0.89, and P < 0.01). However, the PETCO2-PaCO2 gradients (mean 7.0 +/- 3.1 mmHg) were greater than the PTCCO2-PaCO2 gradients (mean 2.3 +/- 2.4 mmHg) at all three levels of minute ventilation (P < 0.05). These differences were greatest when PaCO2 was in the high range (48-60 mmHg). We conclude that the new Fastrac(TM) CO2 monitor is accurate for monitoring carbon dioxide levels during general anaesthesia. The new transcutaneous devices provide an effective method for non-invasive monitoring of CO2 in situations where continuous, precise control of CO2 levels is desired.
引用
收藏
页码:31 / 36
页数:6
相关论文
共 50 条
  • [1] STABILITY OF THE ARTERIAL TO END-TIDAL CARBON-DIOXIDE DIFFERENCE DURING ANESTHESIA FOR PROLONGED NEUROSURGICAL PROCEDURES
    SHARMA, SK
    MCGUIRE, GP
    CRUISE, CJE
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1995, 42 (06): : 498 - 503
  • [2] END-TIDAL CARBON-DIOXIDE MEASUREMENT IN INFANTS AND CHILDREN DURING AND AFTER GENERAL-ANESTHESIA
    CAMPBELL, FA
    MCLEOD, ME
    BISSONNETTE, B
    SWARTZ, JS
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1994, 41 (02): : 107 - 110
  • [3] COMPARISON OF END-TIDAL AND ARTERIAL CARBON-DIOXIDE MEASUREMENTS DURING ANESTHESIA WITH THE LARYNGEAL MASK AIRWAY
    HICKS, IR
    SONI, NC
    SHEPHARD, JN
    BRITISH JOURNAL OF ANAESTHESIA, 1993, 71 (05) : 734 - 735
  • [4] END-TIDAL CARBON-DIOXIDE AS AN INDICATOR OF ARTERIAL CARBON-DIOXIDE IN NEUROINTENSIVE CARE PATIENTS
    RUSSELL, GB
    GRAYBEAL, JM
    JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY, 1992, 4 (04) : 245 - 249
  • [5] STABILITY OF ARTERIAL TO END-TIDAL CARBON-DIOXIDE GRADIENTS DURING POSTOPERATIVE CARDIORESPIRATORY SUPPORT
    RUSSELL, GB
    GRAYBEAL, JM
    STROUT, JC
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1990, 37 (05): : 560 - 566
  • [6] END-TIDAL CARBON-DIOXIDE MEASUREMENT DURING PEDIATRIC ANESTHESIA - THE INFLUENCE OF SITE OF MEASUREMENT AND RESPIRATORY RATE
    FRETSCHNER, R
    WARTH, H
    DEUSCH, H
    KLOSS, T
    ANAESTHESIST, 1992, 41 (08): : 463 - 467
  • [7] Non-invasive carbon dioxide monitoring in patients with cystic fibrosis during general anesthesia: end-tidal versus transcutaneous techniques
    May, Anne
    Humston, Chris
    Rice, Julie
    Nemastil, Christopher J.
    Salvator, Ann
    Tobias, Joseph
    JOURNAL OF ANESTHESIA, 2020, 34 (01) : 66 - 71
  • [8] CORRELATION OF THE PARTIAL PRESSURE OF ARTERIAL CARBON DIOXIDE, END-TIDAL CARBON DIOXIDE AND TRANSCUTANEOUS CARBON DIOXIDE IN PATIENTS WITH RESPIRATORY DISEASE
    Tarke, Chandrakant R.
    Kumar, R.
    Vijayan, V.
    RESPIROLOGY, 2011, 16 : 16 - 17
  • [9] SENSITIVITY OF A DISPOSABLE END-TIDAL CARBON-DIOXIDE DETECTOR
    JONES, BR
    DORSEY, MJ
    JOURNAL OF CLINICAL MONITORING, 1991, 7 (03): : 268 - 270
  • [10] ACCURACY OF END-TIDAL CARBON-DIOXIDE TENSION ANALYZERS
    RAEMER, DB
    CALALANG, I
    JOURNAL OF CLINICAL MONITORING, 1991, 7 (02): : 195 - 208