摘要
The characteristics of PIN diodes have been determined. These diodes have lower leakage currents and noise than other types of Si radiation detectors. The energy resolutions (FWHM) of a 1 cm2 x 0.5 mm PIN diode measured with a pulser, 122.0 keV gamma-rays, 193 keV electrons and 5.5 MeV alpha-particles were 2.6, 2.8, 2.9 and 11.0 keV, respectively. For a 6 mm x 6 mm x 0.2 mm PIN diode, the resolutions (FWHM) for a pulser, 60 keV gamma-rays, 193 keV electrons and 5.5-Mev alpha-particles were 2.1, 2.2, 2.4 and 10.8 keV, respectively.