DETERMINATION OF CHOLESTEROL AND CHOLESTEROL ESTER WITH NOVEL ENZYME MICROSENSORS

被引:83
作者
MOTONAKA, J [1 ]
FAULKNER, LR [1 ]
机构
[1] UNIV ILLINOIS,DEPT CHEM,URBANA,IL 61801
关键词
D O I
10.1021/ac00070a015
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Enzyme microsensors using cholesterol oxidase (EC 1.1.3.6) and cholesterol esterase (EC 3.1.1.13) were developed for measuring cholesterol and cholesterol ester. The platinum microsensors (platinum diameter, 50 mum) were etched in hot aqua regia to create a cavity at their tip. A porous composite material prepared from acetylene black and Teflon emulsion was packed into this cavity and the redox mediator [Os(bpy)3](PF6)2 was monitored by cyclic voltammetry in the potential range of 200-900 mV. The microsensors were dipped overnight in buffer solution containing the desired enzyme to immobilize it on the tip by adsorption. Calibration curves for measurements of cholesterol and cholesterol ester, the effects of pH, temperature, and concomitant compounds, the lifetime of the microsensors, and their availability for measuring cholesterol and cholesterol ester in urine were examined. Under optimal conditions, the response of the sensors was linear in concentration ranges of 5 muM-0.47 mM cholesterol and 2 muM-1.00 mM cholesterol ester.
引用
收藏
页码:3258 / 3261
页数:4
相关论文
共 11 条
[1]   ENZYME ELECTRODE WITH COLLAGEN-IMMOBILIZED CHOLESTEROL OXIDASE FOR THE MICRO-DETERMINATION OF FREE-CHOLESTEROL [J].
BERTRAND, C ;
COULET, PR ;
GAUTHERON, DC .
ANALYTICAL LETTERS PART B-CLINICAL AND BIOCHEMICAL ANALYSIS, 1979, 12 (14) :1477-1488
[2]   ELECTRON HOPPING VS MOLECULAR-DIFFUSION AS CHARGE-TRANSFER MECHANISMS IN REDOX POLYMER-FILMS [J].
BUTTRY, DA ;
ANSON, FC .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1981, 130 (1-3) :333-338
[3]  
CLARK LC, 1975, 3RD INT C ENZ ENG, V3
[4]   A WATER-INSOLUBLE POLYANIONIC DERIVATIVE OF TRYPSIN .2. EFFECT OF POLYELECTROLYTE CARRIER ON KINETIC BEHAVIOR OF BOUND TRYPSIN [J].
GOLDSTEIN, L ;
LEVIN, Y ;
KATCHALSKI, E .
BIOCHEMISTRY, 1964, 3 (12) :1913-+
[5]  
GUILBAULT GG, 1984, ANAL USES IMMOBILIZE, P137
[6]   AMPEROMETRIC ENZYMATIC DETERMINATION OF TOTAL CHOLESTEROL IN HUMAN-SERUM WITH TUBULAR CARBON ELECTRODES [J].
HAHN, Y ;
OLSON, CL .
ANALYTICAL CHEMISTRY, 1979, 51 (03) :444-449
[7]  
ISHIDATE M, 1986, DAI 2 KAISEI NIPPON
[8]   DETERMINATION OF FREE AND TOTAL CHOLESTEROL IN HUMAN BILE SAMPLES USING AN ENZYME ELECTRODE [J].
MASCINI, M ;
TOMASSETTI, M ;
IANNELLO, M .
CLINICA CHIMICA ACTA, 1983, 132 (01) :7-15
[9]   PREPARATION AND PROPERTIES OF A MICRO ENZYME SENSOR FOR CREATINE [J].
MOTONAKA, J ;
TAKABAYASHI, H ;
IKEDA, S ;
TANAKA, N .
ANALYTICAL LETTERS, 1990, 23 (11) :1981-1991
[10]   DIRECT MEASURE OF THE LOW-DENSITY FRACTIONS OF SERUM-CHOLESTEROL [J].
PURDIE, N ;
MURPHY, LH ;
PURDIE, RB .