Establishment of diagnostic reference levels in computed tomography for select procedures in Pudhuchery, India

被引:24
作者
Saravanakumar, A. [1 ]
Vaideki, K. [1 ]
Govindarajan, K. N. [1 ]
Jayakumar, S. [1 ]
机构
[1] PSG Coll Technol, Dept Phys & Mat Sci, Coimbatore 641004, Tamil Nadu, India
关键词
Computed tomography; CTDIw; CTDIv; dose length product; dose reference level; pencil ionization chamber; polymethylmethacrylate phantom;
D O I
10.4103/0971-6203.125509
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Computed tomography (CT) scanner under operating conditions has become a major source of human exposure to diagnostic X-rays. In this context, weighed CT dose index (CTDIw ), volumetric CT dose index (CTDIv ), and dose length product (DLP) are important parameter to assess procedures in CT imaging as surrogate dose quantities for patient dose optimization. The current work aims to estimate the existing dose level of CT scanner for head, chest, and abdomen procedures in Pudhuchery in south India and establish dose reference level (DRL) for the region. The study was carried out for six CT scanners in six different radiology departments using 100 mm long pencil ionization chamber and polymethylmethacrylate (PMMA) phantom. From each CT scanner, data pertaining to patient and machine details were collected for 50 head, 50 chest, and 50 abdomen procedures performed over a period of 1 year. The experimental work was carried out using the machine operating parameters used during the procedures. Initially, dose received in the phantom at the center and periphery was measured by five point method. Using these values CTDIw , CTDIv , and DLP were calculated. The DRL is established based on the third quartile value of CTDIv and DLP which is 32 mGy and 925 mGy.cm for head, 12 mGy and 456 mGy.cm for chest, and 16 mGy and 482 mGy.cm for abdomen procedures. These values are well below European Commission Dose Reference Level (EC DRL) and comparable with the third quartile value reported for Tamil Nadu region in India. The present study is the first of its kind to determine the DRL for scanners operating in the Pudhuchery region. Similar studies in other regions of India are necessary in order to establish a National Dose Reference Level.
引用
收藏
页码:50 / 55
页数:6
相关论文
共 13 条
[1]  
[Anonymous], 2006, HLTH RISKS EXP LOW L
[2]  
European Commission, 16262 EUR EN EUR COM
[3]   Cancer risks from diagnostic radiology [J].
Hall, E. J. ;
Brenner, D. J. .
BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (965) :362-378
[4]   A contribution to the establishment of diagnostic reference levels in CT [J].
Hatziioannou, K ;
Papanastassiou, E ;
Delichas, M ;
Bousbouras, P .
BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (908) :541-545
[5]   Overview of patient dosimetry in diagnostic radiology in the USA for the past 50 years [J].
Huda, Walter ;
Nickoloff, Edward L. ;
Boone, John M. .
MEDICAL PHYSICS, 2008, 35 (12) :5713-5728
[6]  
International Atomic Energy Agency, 2004, IAEATECDOC1423
[7]  
International Atomic Energy Agency, 1996, IAEA SAF SER, V115
[8]   Dose reduction in computed tomography by individualized scan protocols [J].
Jangland, L ;
Sanner, E ;
Persliden, J .
ACTA RADIOLOGICA, 2004, 45 (03) :301-307
[9]   Strategies for CT radiation dose optimization [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Hamberg, LM ;
Blake, MA ;
Shephard, JA ;
Saini, S .
RADIOLOGY, 2004, 230 (03) :619-628
[10]   Radiation safety concerns and diagnostic reference levels for computed tomography scanners in Tamil Nadu [J].
Livingstone, Roshan S. ;
Dinakaran, Paul M. .
JOURNAL OF MEDICAL PHYSICS, 2011, 36 (01) :40-45