A 12 DETECTOR COINCIDENCE SPECTROMETER FOR ONLINE AND OFF-LINE MEASUREMENTS OF NUCLEAR G-FACTORS

被引:15
作者
ALFTER, I
BODENSTEDT, E
HAMER, B
VANDENHOFF, J
KNICHEL, W
MUNNING, H
PIEL, S
SCHUTH, J
SAJOK, R
机构
[1] Institut für Strahlen-, Kernphysik der Universität Bonn
关键词
D O I
10.1016/0168-9002(92)90061-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A twelve detector gamma - gamma directional correlation setup is described which has been designed and constructed for high precision on-line and off-line IPAC measurements. A ring of twelve BaF2 detectors which may be replaced by Ge detectors is arranged around a superconducting magnet providing a maximum magnetic field of 4.5 T. In on-line experiments the magnetic field has the direction of the beam which passes through a central bore of the magnet. The data acquisition system stores directly the E(gamma1) - E(gamma2) coincidence matrices for all 66 two-detector combinations in two 8 Mbyte RAMs, one for each of the two field directions. This allows a very fast data evaluation with a small on-line computer. The apparatus got into operation one-and-a-halve years ago and revealed excellent properties.
引用
收藏
页码:506 / 520
页数:15
相关论文
共 21 条
[1]  
BECK FA, 1991, EUROPHYS NEWS, V22, P97
[2]   MEASUREMENT OF THE G-FACTOR OF THE 2(1)+ STATE OF PT-192 IN AN EXTERNAL MAGNETIC-FIELD - 1ST EXPERIMENTS WITH A NEW 12 DETECTOR IPAC APPARATUS [J].
BODENSTEDT, E ;
HAMER, B ;
HERZOG, P ;
VANDENHOFF, J ;
MUNNING, H ;
PIEL, S ;
PRINZ, J ;
SAJOK, R .
ZEITSCHRIFT FUR PHYSIK A-HADRONS AND NUCLEI, 1992, 342 (03) :249-256
[3]   FAST TIMING WITH PHOTOMULTIPLIER DYNODE PULSES [J].
DEVRIES, J ;
KELLING, FET .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 262 (2-3) :385-393
[4]   THE DATA ACQUISITION-SYSTEM FOR THE CRYSTAL BALL AT LNS [J].
FINOCCHIARO, P ;
MAIOLINO, C ;
PIATTELLI, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 271 (03) :557-562
[5]   56-CHANNEL GONIOMETER FOR GAMMA-GAMMA DIRECTIONAL CORRELATION STUDIES [J].
GERHOLM, TR ;
GIDEFELDT, L ;
ERIKSSON, L ;
CHO, ZH ;
PETTERSSON, BG .
NUCLEAR INSTRUMENTS & METHODS, 1972, 100 (01) :33-+
[6]  
HASHIMOTO T, 1971, RADIOANALYTICAL CHEM, V9, P251
[7]  
HUBEL H, ISKP198384 U BONN
[8]   DESIGN OF A BISMUTH-GERMANATE ANTI-COMPTON SPECTROMETER AND ITS USE IN NUCLEAR SPECTROSCOPY [J].
LIEDER, RM ;
JAGER, H ;
NESKAKIS, A ;
VENKOVA, T ;
MICHEL, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 220 (2-3) :363-370
[9]   DATA ACQUISITION-SYSTEM FOR THE 8-PI SPECTROMETER - A MULTIDETECTOR ARRAY FOR GAMMA-RAY SPECTROSCOPY [J].
MARTIN, JP ;
RADFORD, DC ;
BEAULIEU, M ;
TARAS, P ;
WARD, D ;
ANDREWS, HR ;
AYOTTE, G ;
SHARP, FJ ;
WADDINGTON, JC ;
HAUSSER, O ;
GASCON, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 257 (02) :301-308
[10]   PROPERTIES OF BAF2 SCINTILLATORS IN CHARGED-PARTICLE DETECTION [J].
MURAKAMI, T ;
KASAGI, J ;
TACHIBANAKI, H ;
YOSHIDA, K ;
SHIBATA, Y ;
NAKAGAWA, T ;
OGIHARA, M ;
LEE, SM ;
KUBO, T ;
MOTOBAYASHI, T .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 253 (01) :163-165