Electrospray ionization-voltage sweep-Ion mobility spectrometry for biomolecules and complex samples

被引:3
作者
Davis, Eric J. [1 ]
Grows, Kristopher F. [2 ]
Siems, William F. [2 ]
Hill, Herbert H., Jr. [2 ]
机构
[1] Fresno Pacific Univ, Sch Nat Sci, 1717 S Chestnut Ave, Fresno, CA 93702 USA
[2] Washington State Univ, Dept Chem, Pullman, WA 99164 USA
关键词
Voltage sweep; Voltage-sweep; VSIMS; IMS; ESI; ESI-IMS; ESI-VSIMS; Electrospray; DTIMS; Ion mobility spectrometry; Peak capacity; Resolving power; Resolution;
D O I
10.1007/s12127-014-0155-7
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Voltage Sweep Ion Mobility Spectrometry (VSIMS) has been applied to complex samples using electrospray ionization (ESI). The usable range of VSIMS has been extended from that obtained in previous studies where only volatile compounds were investigated. Using ESI, VSIMS was evaluated with compounds with reduced mobility values as low as 0.3 V(2)cm(-1) s(-1). The primary advantage of VSIMS is to enable a drift time ion mobility spectrometer (DTIMS) to detect both fast and slow moving ions at optimal resolving power, thus improving the peak capacity. In this work ESI-VSIMS was applied to a series of small peptides and drugs spanning a large range of reduced mobility values in order to demonstrate ESI-VSIMS to separation. To demonstrate improved peak capacity of IMS with voltage scan operation, oligomers of silicone oil provided a series of evenly-spaced peaks, ranging in reduced mobility values from 0.85 to 0.3 V(2)cm(-1) s(-1). The peak capacity of 61 for a standard IMS was improved to 102 when voltage sweep operation was employed. In addition, VSIMS increased the average resolving power of the DTIMS from 66 to 106 for silicone oil.
引用
收藏
页码:147 / 156
页数:10
相关论文
共 23 条
[11]  
EICEMAN GA, 1994, ION MOBILITY SPECTRO
[12]   A critical review of ion mobility spectrometry for the detection of explosives and explosive related compounds [J].
Ewing, RG ;
Atkinson, DA ;
Eiceman, GA ;
Ewing, GJ .
TALANTA, 2001, 54 (03) :515-529
[13]   A novel non-radioactive electron source for ion mobility spectrometry [J].
Gunzer, Frank ;
Ulrich, Andreas ;
Baether, Wolfgang .
INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2010, 13 (01) :9-16
[14]   Predicting optimal resolving power for ambient pressure ion mobility spectrometry [J].
Kanu, Abu B. ;
Gribb, Molly M. ;
Hill, Herbert H., Jr. .
ANALYTICAL CHEMISTRY, 2008, 80 (17) :6610-6619
[15]   Identity confirmation of drugs and explosives in ion mobility spectrometry using a secondary drift gas [J].
Kanu, Abu B. ;
Hill, Herbert H., Jr. .
TALANTA, 2007, 73 (04) :692-699
[16]   Results of a transient simulation of a drift tube ion mobility spectrometer considering charge repulsion, ion loss at metallic surfaces and ion generation [J].
Langejuergen, Jens ;
Cochems, Philipp ;
Zimmermann, Stefan .
INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2012, 15 (04) :247-255
[17]  
Nogare S. D., 1958, ANAL CHEM, V30, P1157
[18]   An investigation of the mobility separation of some peptide and protein ions using a new hybrid quadrupole/travelling wave IMS/oa-ToF instrument [J].
Pringle, Steven D. ;
Giles, Kevin ;
Wildgoose, Jason L. ;
Williams, Jonathan P. ;
Slade, Susan E. ;
Thalassinos, Konstantinos ;
Bateman, Robert H. ;
Bowers, Michael T. ;
Scrivens, James H. .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2007, 261 (01) :1-12
[19]   Travelling wave ion mobility mass spectrometry studies of protein structure: biological significance and comparison with X-ray crystallography and nuclear magnetic resonance spectroscopy measurements [J].
Scarff, Charlotte A. ;
Thalassinos, Konstantinos ;
Hilton, Gillian R. ;
Scrivens, James H. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2008, 22 (20) :3297-3304
[20]   MEASURING THE RESOLVING POWER OF ION MOBILITY SPECTROMETERS [J].
SIEMS, WF ;
WU, C ;
TARVER, EE ;
HILL, HH ;
LARSEN, PR ;
MCMINN, DG .
ANALYTICAL CHEMISTRY, 1994, 66 (23) :4195-4201