Electrospray ionization-voltage sweep-Ion mobility spectrometry for biomolecules and complex samples

被引:3
作者
Davis, Eric J. [1 ]
Grows, Kristopher F. [2 ]
Siems, William F. [2 ]
Hill, Herbert H., Jr. [2 ]
机构
[1] Fresno Pacific Univ, Sch Nat Sci, 1717 S Chestnut Ave, Fresno, CA 93702 USA
[2] Washington State Univ, Dept Chem, Pullman, WA 99164 USA
关键词
Voltage sweep; Voltage-sweep; VSIMS; IMS; ESI; ESI-IMS; ESI-VSIMS; Electrospray; DTIMS; Ion mobility spectrometry; Peak capacity; Resolving power; Resolution;
D O I
10.1007/s12127-014-0155-7
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Voltage Sweep Ion Mobility Spectrometry (VSIMS) has been applied to complex samples using electrospray ionization (ESI). The usable range of VSIMS has been extended from that obtained in previous studies where only volatile compounds were investigated. Using ESI, VSIMS was evaluated with compounds with reduced mobility values as low as 0.3 V(2)cm(-1) s(-1). The primary advantage of VSIMS is to enable a drift time ion mobility spectrometer (DTIMS) to detect both fast and slow moving ions at optimal resolving power, thus improving the peak capacity. In this work ESI-VSIMS was applied to a series of small peptides and drugs spanning a large range of reduced mobility values in order to demonstrate ESI-VSIMS to separation. To demonstrate improved peak capacity of IMS with voltage scan operation, oligomers of silicone oil provided a series of evenly-spaced peaks, ranging in reduced mobility values from 0.85 to 0.3 V(2)cm(-1) s(-1). The peak capacity of 61 for a standard IMS was improved to 102 when voltage sweep operation was employed. In addition, VSIMS increased the average resolving power of the DTIMS from 66 to 106 for silicone oil.
引用
收藏
页码:147 / 156
页数:10
相关论文
共 23 条
  • [1] Performance evaluation of a miniature ion mobility spectrometer drift cell for application in hand-held explosives detection ion mobility spectrometers
    Babis, J. S.
    Sperline, R. P.
    Knight, A. K.
    Jones, D. A.
    Gresham, C. A.
    Denton, M. B.
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 395 (02) : 411 - 419
  • [2] NAKED PROTEIN CONFORMATIONS - CYTOCHROME-C IN THE GAS-PHASE
    CLEMMER, DE
    HUDGINS, RR
    JARROLD, MF
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1995, 117 (40) : 10141 - 10142
  • [3] Clowers BH, 2005, THESIS
  • [4] The role of oxygen in the formation of TNT product ions in ion mobility spectrometry
    Daum, KA
    Atkinson, DA
    Ewing, RG
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 214 (02) : 257 - 267
  • [5] Voltage Sweep Ion Mobility Spectrometry
    Davis, Eric J.
    Williams, Michael D.
    Siems, William F.
    Hill, Herbert H., Jr.
    [J]. ANALYTICAL CHEMISTRY, 2011, 83 (04) : 1260 - 1267
  • [6] High-Pressure Ion Mobility Spectrometry
    Davis, Eric J.
    Dwivedi, Prabha
    Tam, Maggie
    Siems, William F.
    Hill, Herbert H.
    [J]. ANALYTICAL CHEMISTRY, 2009, 81 (09) : 3270 - 3275
  • [7] Comprehensive software suite for the operation, maintenance, and evaluation of an ion mobility spectrometer
    Davis, Eric James
    Clowers, Brian H.
    Siems, William F.
    Hill, Herbert H., Jr.
    [J]. INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2011, 14 (2-3) : 117 - 124
  • [8] Rapid resolution of carbohydrate isomers by electrospray ionization ambient pressure ion mobility spectrometry-time-of-flight mass spectrometry (ESI-APIMS-TOFMS)
    Dwivedi, Prabha
    Bendiak, Brad
    Clowers, Brian H.
    Hill, Herbert H., Jr.
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (07) : 1163 - 1175
  • [9] Ion mobility spectrometers in national defense.
    Eiceman, GA
    Stone, JA
    [J]. ANALYTICAL CHEMISTRY, 2004, 76 (21) : 390A - 397A
  • [10] Chemical standards in ion mobility spectrometry
    Eiceman, GA
    Nazarov, EG
    Stone, JA
    [J]. ANALYTICA CHIMICA ACTA, 2003, 493 (02) : 185 - 194