REMOVAL OF RETAINED INTRACORONARY PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY EQUIPMENT BY A PERCUTANEOUS TWIN GUIDEWIRE METHOD

被引:29
作者
GURLEY, JC
BOOTH, DC
HIXSON, C
SMITH, MD
机构
[1] UNIV KENTUCKY,MED CTR,DIV CARDIOVASC MED,LEXINGTON,KY 40536
[2] VET ADM MED CTR,DIV CARDIOVASC MED,LEXINGTON,KY 40511
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1990年 / 19卷 / 04期
关键词
balloon rupture; catheter retrieval methods; complications;
D O I
10.1002/ccd.1810190408
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Balloon rupture during percutaneous transluminal coronary angioplasty (PTCA) resulted in embolization of a guidewire‐balloon fragment into the distal circumflex artery. Access to the circumflex was technically difficult because of aortic root dilation. The fragment was removed by inserting two PTCA guidewires into the artery and twisting them together to produce a helix. The helix propagated distally, entrapping the embolized fragment for removal. This report describes an unusual mode of failure for a balloon‐ona‐ wire dilating catheter and a new percutaneous removal technique that may be applied to retained catheter debris in small, inaccessible vessels. Copyright © 1990 John Wiley & Sons, Ltd.
引用
收藏
页码:251 / 256
页数:6
相关论文
共 11 条
[1]  
ARCEGONZOLEZ JM, 1988, J AM COL CARDIOL, V10, P218
[2]   SURGICAL REMOVAL OF THE INTRACORONARY PORTION OF A FRACTURED ANGIOPLASTY GUIDEWIRE [J].
BACHENHEIMER, LC ;
GREEN, CE ;
ROSING, DR ;
WALLACE, RB .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (11) :946-947
[3]   RETAINED PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY EQUIPMENT COMPONENTS AND THEIR MANAGEMENT [J].
HARTZLER, GO ;
RUTHERFORD, BD ;
MCCONAHAY, DR .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 60 (16) :1260-1264
[4]  
KOHNSARI S, 1986, AM J CARDIOL, V56, P855
[5]   SUCCESSFUL PERCUTANEOUS REMOVAL OF RETAINED BROKEN CORONARY ANGIOPLASTY GUIDEWIRE [J].
KRONE, RJ .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1986, 12 (06) :409-410
[6]   GUIDE WIRE ENTRAPMENT DURING PTCA - A POTENTIALLY DANGEROUS COMPLICATION [J].
LOTAN, C ;
HASIN, Y ;
STONE, D ;
MEYERS, S ;
APPLEBAUM, A ;
GOTSMAN, MS .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1987, 13 (05) :309-312
[7]   TRANSCATHETER RETRIEVAL OF INTRACORONARY DETACHED ANGIOPLASTY GUIDEWIRE SEGMENT [J].
MIKOLICH, JR ;
HANSON, MW .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1988, 15 (01) :44-46
[8]   INTRAVASCULAR RETRIEVAL OF BROKEN GUIDE WIRE FROM THE ASCENDING AORTA AFTER PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
STEELE, PM ;
HOLMES, DR ;
MANKIN, HT ;
SCHAFF, HV .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1985, 11 (06) :623-628
[9]   FRACTURE OF A BALLOON ON A WIRE DEVICE DURING CORONARY ANGIOPLASTY [J].
VANDENBRAND, M ;
DEFEYTER, P ;
SERRUYS, P ;
ZIJLSTRA, F ;
BOS, E .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1989, 16 (04) :253-257
[10]  
VROLIX M, 1988, CATHETER CARDIO DIAG, V15, P99