Uncertainty in positioning ion chamber at reference depth for various water phantoms

被引:7
作者
Kinoshita, Naoki [1 ,2 ]
Oguchi, Hiroshi [1 ]
Adachi, Toshiki [2 ]
Shioura, Hiroki [3 ]
Kimura, Hirohiko [3 ]
机构
[1] Nagoya Univ, Grad Sch Med, Dept Radiol & Med Lab Sci, Nagoya, Aichi 4618673, Japan
[2] Univ Fukui Hosp, Radiol Ctr, Eiheiji, Fukui 9101193, Japan
[3] Univ Fukui Hosp, Dept Radiol, Eiheiji, Fukui 9101193, Japan
关键词
Positioning uncertainty; Ionization chamber; Water phantom; External photon beam radiotherapy; Calibration;
D O I
10.1016/j.rpor.2018.03.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Uncertainty in the calibration of high-energy radiation sources is dependent on user and equipment type. Aim: We evaluated the uncertainty in the positioning of a cylindrical chamber at a reference depth for reference dosimetry of high-energy photon beams and the resulting uncertainty in the chamber readings for 6- and 10-MV photon beams. The aim was to investigate major contributions to the positioning uncertainty to reduce the uncertainty in calibration for external photon beam radiotherapy. Materials and methods: The following phantoms were used: Dose View 1D, WP1D, 1D SCAN-NER, and QWP-07 as one-dimensional (1D) phantoms for a vertical-beam geometry; GRI-7632 as a phantom for a fixed waterproofing sleeve; and PTW type 41023 and QWP-04 as 1D phantoms for a horizontal-beam geometry. The uncertainties were analyzed as per the Guide to the Expression of Uncertainty in Measurement. Results: The positioning and resultant uncertainties in chamber readings ranged from 0.22 to 0.35 mm and 0.12-0.25%, respectively, among the phantoms (using a coverage factor k = 1 in both cases). The major contributions to positioning uncertainty are: definition of the origin for phantoms among users for the 1D phantoms for a vertical-beam geometry, water level adjustment among users for the phantom for a fixed waterproofing sleeve, phantom window deformation, and non-water material of the window for the 1D phantoms for a horizontal-beam geometry. Conclusion: The positioning and resultant uncertainties in chamber readings exhibited minor differences among the seven phantoms. The major components of these uncertainties differed among the phantom types investigated. (c) 2018 Greater Poland Cancer Centre. Published by Elsevier Sp.zo.o. All rights reserved.
引用
收藏
页码:199 / 206
页数:8
相关论文
共 15 条
  • [1] AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    Almond, PR
    Biggs, PJ
    Coursey, BM
    Hanson, WF
    Huq, MS
    Nath, R
    Rogers, DWO
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1847 - 1870
  • [2] Andreo P, 2000, TECHNICAL REPORTS SE, V398
  • [3] Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water
    Arib, Mehenna
    Medjadj, Toufik
    Boudouma, Youcef
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (03): : 55 - 64
  • [4] BIPM IEC IFCC ISO IUPAC IUPAP OIML, 2008, 1 JCGMWG
  • [5] Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration
    Castro, Pablo
    Garcia-Vicente, Feliciano
    Minguez, Cristina
    Floriano, Alejandro
    Sevillano, David
    Perez, Leopoldo
    Torres, Juan J.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (01): : 70 - 86
  • [6] Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM
    Das, Indra J.
    Cheng, Chee-Wai
    Watts, Ronald J.
    Ahnesjo, Anders
    Gibbons, John
    Li, X. Allen
    Lowenstein, Jessica
    Mitra, Raj K.
    Simon, William E.
    Zhu, Timothy C.
    [J]. MEDICAL PHYSICS, 2008, 35 (09) : 4186 - 4215
  • [7] IAEA. (International Atomic Energy Agency), 2008, TECDOC1585 IAEA
  • [8] Japan Society of Medical Physics, 2012, STAND DOS ABS DOS WA
  • [9] McEwen M, 2014, MED PHYS, V41
  • [10] The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams
    McEwen, M. R.
    Kawrakow, I.
    Ross, C. K.
    [J]. MEDICAL PHYSICS, 2008, 35 (03) : 950 - 958