AN IMPROVED MICRO-EXTRACTION CELL FOR SUPERCRITICAL FLUID EXTRACTION AND CHROMATOGRAPHY OF FATTY-ACIDS

被引:8
|
作者
ARTZ, WE
SAUER, RM
机构
[1] University of Illinois, Urbana, 61801, IL
关键词
FATTY ACIDS; SUPERCRITICAL FLUID CHROMATOGRAPHY; SUPERCRITICAL FLUID EXTRACTION;
D O I
10.1007/BF02636056
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
An improved supercritical fluid micro-extraction cell of increased reliability was designed for on-line supercritical fluid extraction and chromatography (SFE/SFC) of food and other lipid-related samples. The key components in the modified cell include a Swagelok stainless steel reducing union with a dual ferrule as the cell with polyetheretherketone (PEEK) ferrules and nuts to connect the cell to the control valve. The new cell did not leak under all conditions examined (100-500 atmospheres, 40-80-degrees-C), even after numerous extractions (> 250). The quantitative performance of the cell was evaluated with fatty acid standard solutions, technical grade fatty acid sources and wheat flour. The percent relative error (%RE) for the fatty acid standards and technical-grade fatty acid samples was less-than-or-equal-to 6.0% for oleic, linoleic and linolenic acid. The % RE for oleic and linoleic acid in the whole-wheat samples was less-than-or-equal-to 10%. The results demonstrate that the new extraction cell can be used for quantitative extractions and that the sensitivity of the SFE/SFC technique is excellent. Similar SFE/SFC methods could prove useful in studying the interaction of free fatty acids with various food components such as enzymes, amylose and proteins.
引用
收藏
页码:309 / 313
页数:5
相关论文
共 50 条