Accuracy of a real-time continuous glucose monitoring system in children with septic shock: A pilot study

被引:9
|
作者
Prabhudesai, Sumant [1 ]
Kanjani, Amruta [2 ]
Bhagat, Isha [3 ]
Ravikumar, Karnam G. [4 ]
Ramachandran, Bala [4 ]
机构
[1] Royal Manchester Childrens Hosp, Oxford Rd, Manchester M13 9WL, Lancs, England
[2] Columbia Asia Hosp, Ahmedebad, Gujrat, India
[3] PD Hinduja Natl Hosp & Med Res Ctr, Bombay, Maharashtra, India
[4] Kanchi Kamakoti CHILDS Trust Hosp, Madras, Tamil Nadu, India
关键词
Children; continuous glucose monitoring system; septic shock;
D O I
10.4103/0972-5229.169331
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The aim of this prospective, observational study was to determine the accuracy of a real-time continuous glucose monitoring system (CGMS) in children with septic shock. Subjects and Methods: Children aged 30 days to 18 years admitted to the Pediatric Intensive Care Unit with septic shock were included. A real-time CGMS sensor was used to obtain interstitial glucose readings. CGMS readings were compared statistically with simultaneous laboratory blood glucose (BG). Results: Nineteen children were included, and 235 pairs of BG-CGMS readings were obtained. BG and CGMS had a correlation coefficient of 0.61 (P < 0.001) and a median relative absolute difference of 17.29%. On Clarke's error grid analysis, 222 ( 94.5%) readings were in the clinically acceptable zones (A and B). When BG was < 70, 70-180, and > 180 mg/dL, 44%, 100%, and 76.9% readings were in zones A and B, respectively (P < 0.001). The accuracy of CGMS was not affected by the presence of edema, acidosis, vasopressors, steroids, or renal replacement therapy. On receiver operating characteristics curve analysis, a CGMS reading < 97 mg/dL predicted hypoglycemia (sensitivity 85.2%, specificity 75%, area under the curve [AUC] =0.85). A reading > 141 mg/dL predicted hyperglycemia (sensitivity 84.6%, specificity 89.6%, AUC =0.87). Conclusion: CGMS provides a fairly, accurate estimate of BG in children with septic shock. It is unaffected by a variety of clinical variables. The accuracy over extremes of blood sugar may be a concern. We recommend larger studies to evaluate its use for the early detection of hypoglycemia and hyperglycemia.
引用
收藏
页码:642 / 647
页数:6
相关论文
共 50 条
  • [1] Real-Time Continuous Glucose Monitoring in an Intensive Care Unit: Better Accuracy in Patients with Septic Shock
    Lorencio, Carol
    Leal, Yenny
    Bonet, Alfonso
    Bondia, Jorge
    Palerm, Cesar C.
    Tache, Abdo
    Sirvent, Josep-Maria
    Vehi, Josep
    DIABETES TECHNOLOGY & THERAPEUTICS, 2012, 14 (07) : 568 - 575
  • [2] Accuracy of real-time continuous glucose monitoring in the MiniMed paradigm system
    Mastrototaro, John
    Soundararajan, Gopikrishnan
    Cooper, Ken
    Shah, Rajiv
    DIABETES, 2007, 56 : A112 - A112
  • [3] PERFORMANCE OF A NEW REAL-TIME CONTINUOUS GLUCOSE MONITORING SYSTEM: A MULTICENTER PILOT STUDY
    Zhou, J.
    Zhang, S.
    Li, L.
    Wang, Y.
    Lu, W.
    Sheng, C.
    Li, Y.
    Bao, Y.
    Jia, W.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2018, 20 : A72 - A73
  • [4] Performance of a new real-time continuous glucose monitoring system: A multicenter pilot study
    Zhou, Jian
    Zhang, Shuo
    Li, Liang
    Wang, Yufei
    Lu, Wei
    Sheng, Chunjun
    Li, Yiming
    Bao, Yuqian
    Jia, Weiping
    JOURNAL OF DIABETES INVESTIGATION, 2018, 9 (02): : 286 - 293
  • [5] Real-Time Improvement of Continuous Glucose Monitoring Accuracy
    Facchinetti, Andrea
    Sparacino, Giovanni
    Guerra, Stefania
    Luijf, Yoeri M.
    DeVries, J. Hans
    Mader, Julia K.
    Ellmerer, Martin
    Benesch, Carsten
    Heinemann, Lutz
    Bruttomesso, Daniela
    Avogaro, Angelo
    Cobelli, Claudio
    DIABETES CARE, 2013, 36 (04) : 793 - 800
  • [6] Accuracy of real-time glucose values using continuous glucose monitoring:: the Guardian™ II continuous glucose monitoring system
    Schwartz, SL
    Buchea, O
    Shin, J
    Hoss, U
    Shah, R
    DIABETOLOGIA, 2004, 47 : A315 - A315
  • [7] Accuracy of a new real-time continuous glucose monitoring algorithm
    Navas de Solis, M. S.
    AVANCES EN DIABETOLOGIA, 2010, 26 (02): : 129 - 130
  • [8] Pre- and Postoperative Accuracy and Safety of a Real-Time Continuous Glucose Monitoring System in Cardiac Surgical Patients: A Randomized Pilot Study
    Logtenberg, Susan J.
    Kleefstra, Nanne
    Snellen, Ferdinand T.
    Groenier, Klaas H.
    Slingerland, Robbert J.
    Nierich, Arno P.
    Bilo, Henk J.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 (01) : 31 - 37
  • [9] Feasibility of Real-Time Continuous Glucose Monitoring Telemetry System in an Inpatient Diabetes Unit: A Pilot Study
    Dillmann, Clementine
    Amoura, Lamia
    Mostaine, Fatema Fall
    Coste, Adrien
    Bounyar, Leila
    Kessler, Laurence
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2022, 16 (04): : 955 - 961
  • [10] Assessment of the accuracy of real-time continuous glucose monitoring system and its correlated factors
    洛佩
    ChinaMedicalAbstracts(InternalMedicine), 2014, 31 (01) : 19 - 19