THIN-FILM HIGH-TEMPERATURE SUPERCONDUCTING RF COILS FOR LOW-FIELD MRI

被引:51
作者
VANHETEREN, JG
JAMES, TW
BOURNE, LC
机构
[1] TOSHIBA AMERICA MRI INC,S SAN FRANCISCO,CA
[2] SUPERCONDUCTOR TECHNOL INC,SANTA BARBARA,CA 93111
关键词
RF COILS; SUPERCONDUCTORS;
D O I
10.1002/mrm.1910320315
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have built and tested a thin-film high temperature superconducting (HTS) surface coil for MRI at 0.064T. When placed as close as possible to a conductive sample, the 6.7-cm outer diameter HTS coil had a measured signal-to-noise ratio (SNR) 1.8 times higher than a room temperature copper coil of similar size and location. Our results predict that in some cases, SNR gains of about a factor of 2 can be attained in low-field MRI systems by substituting HTS surface coils for copper coils.
引用
收藏
页码:396 / 400
页数:5
相关论文
共 14 条
[1]   A HIGH-TEMPERATURE SUPERCONDUCTING RECEIVER FOR NUCLEAR-MAGNETIC-RESONANCE MICROSCOPY [J].
BLACK, RD ;
EARLY, TA ;
ROEMER, PB ;
MUELLER, OM ;
MOGROCAMPERO, A ;
TURNER, LG ;
JOHNSON, GA .
SCIENCE, 1993, 259 (5096) :793-795
[2]   TRANSPORT CHARACTERISTICS OF TL2BA2CACU2O8 BICRYSTAL GRAIN-BOUNDARY JUNCTIONS AT 77-K [J].
CARDONA, AH ;
SUZUKI, H ;
YAMASHITA, T ;
YOUNG, KH ;
BOURNE, LC .
APPLIED PHYSICS LETTERS, 1993, 62 (04) :411-413
[3]   AN INDUCTIVELY COUPLED, SERIES-TUNED NMR PROBE [J].
DECORPS, M ;
BLONDET, P ;
REUTENAUER, H ;
ALBRAND, JP ;
REMY, C .
JOURNAL OF MAGNETIC RESONANCE, 1985, 65 (01) :100-109
[4]   THE INTRINSIC SIGNAL-TO-NOISE RATIO IN NMR IMAGING [J].
EDELSTEIN, WA ;
GLOVER, GH ;
HARDY, CJ ;
REDINGTON, RW .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (04) :604-618
[5]   INVESTIGATION OF A WHOLE-BODY RECEIVER COIL OPERATING AT LIQUID-NITROGEN TEMPERATURES [J].
HALL, AS ;
BARNARD, B ;
MCARTHUR, P ;
GILDERDALE, DJ ;
YOUNG, IR ;
BYDDER, GM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (02) :230-235
[6]   USE OF HIGH-TEMPERATURE SUPERCONDUCTOR IN A RECEIVER COIL FOR MAGNETIC-RESONANCE-IMAGING [J].
HALL, AS ;
ALFORD, NM ;
BUTTON, TW ;
GILDERDALE, DJ ;
GEHRING, KA ;
YOUNG, IR .
MAGNETIC RESONANCE IN MEDICINE, 1991, 20 (02) :340-343
[7]   EPITAXIAL TL2CABA2CU2O8 THIN-FILMS WITH LOW 9.6-GHZ SURFACE-RESISTANCE AT HIGH-POWER AND ABOVE 77-K [J].
HAMMOND, RB ;
NEGRETE, GV ;
BOURNE, LC ;
STROTHER, DD ;
CARDONA, AH ;
EDDY, MM .
APPLIED PHYSICS LETTERS, 1990, 57 (08) :825-827
[8]   SENSITIVITY OF THE ZEUGMATOGRAPHIC EXPERIMENT INVOLVING HUMAN SAMPLES [J].
HOULT, DI ;
LAUTERBUR, PC .
JOURNAL OF MAGNETIC RESONANCE, 1979, 34 (02) :425-433
[9]  
Jackson J D, 1975, CLASSICAL ELECTRODYN, P298
[10]  
KAUFMAN L, 1989, Magnetic Resonance Quarterly, V5, P283