RESPIRATION MEASUREMENTS IN SMALL-SCALE

被引:15
作者
RASMUSSEN, HN
RASMUSSEN, UF
机构
[1] Laboratory of Biological Chemistry A, August Krogh Institute, University of Copenhagen, Copenhagen
关键词
D O I
10.1006/abio.1993.1040
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The design and operation of a small (<40 μl) reaction vessel with membrane-covered oxygen electrode is described. The equipment's productivity and experimental versatility is similar to that of equipment using 100 times more biological material. Control experiments showed that respiration experiments are made with virtually no systematic errors and random errors of less than 2%. © 1994 Academic Press, Inc. All rights reserved.
引用
收藏
页码:244 / 248
页数:5
相关论文
共 9 条
[1]  
FATT I, 1976, POLAROGRAPHIC OXYGEN
[2]   THE DESIGN AND CONSTRUCTION OF AN OXYGEN MICRO-ELECTRODE FOR MONITORING 20-100-MU-1 VOLUMES [J].
HUSKISSON, NS ;
WARD, PFV ;
MUNN, EA .
BIOCHEMICAL JOURNAL, 1984, 221 (03) :753-758
[3]  
Lessler M A, 1982, Methods Biochem Anal, V28, P175, DOI 10.1002/9780470110485.ch4
[4]  
Linek V., 1988, MEASUREMENT OXYGEN M
[5]   TECHNIQUES FOR BIOCHEMICAL RESPIRATION MEASUREMENTS [J].
RASMUSSEN, HN .
ANALYTICAL BIOCHEMISTRY, 1979, 95 (02) :416-428
[6]   3 FUNCTIONALLY DIFFERENT CYTOCHROME-B REDOX CENTERS IN PIGEON HEART-MITOCHONDRIA [J].
RASMUSSEN, UF ;
RASMUSSEN, HN ;
JORGENSEN, BM .
BIOCHEMICAL JOURNAL, 1982, 201 (02) :311-320
[7]  
RICKWOOD D, 1987, MITOCHONDRIA PRACTIC, P12
[8]  
[No title captured]
[9]  
[No title captured]