CAVITY RING-DOWN TECHNIQUE AND ITS APPLICATION TO THE MEASUREMENT OF ULTRASLOW VELOCITIES

被引:57
作者
AN, KW
YANG, CH
DASARI, EE
FELD, MS
机构
[1] George R. Harrison Spectroscopy Laboratory, Massachusetts Institute of Technology, Cambridge, MA
关键词
D O I
10.1364/OL.20.001068
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have developed a new ring-down technique that does not require a shutter to turn a probe laser on and off. With a rapid cavity scan we can measure a simple exponential cavity decay from which a cavity finesse can be found. When the cavity is scanned slowly, the cavity decay exhibits an amplitude modulation, and an analytic expression is derived for this modulation. With this new technique we measured the ultraslow relative velocity of the mirrors (of the order of micrometers per second) as well as the linewidth (similar to 100 kHz) of the probe laser.
引用
收藏
页码:1068 / 1070
页数:3
相关论文
共 5 条
[1]   MICROLASER - A LASER WITH ONE-ATOM IN AN OPTICAL-RESONATOR [J].
AN, K ;
CHILDS, JJ ;
DASARI, RR ;
FELD, MS .
PHYSICAL REVIEW LETTERS, 1994, 73 (25) :3375-3378
[2]   GF VALUE OF INTERCOMBINATION LINE OF BARIUM, BA1 [J].
PARKINSON, WH ;
TOMKINS, FS .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1978, 68 (04) :535-535
[3]   MEASUREMENT OF ULTRALOW LOSSES IN AN OPTICAL INTERFEROMETER [J].
REMPE, G ;
THOMPSON, RJ ;
KIMBLE, HJ ;
LALEZARI, R .
OPTICS LETTERS, 1992, 17 (05) :363-365
[4]  
SHORE BW, 1990, THEORY COHERENT ATOM, V2, P1545
[5]  
THOMPSON RJ, 1992, PHYS REV LETT, V66, P1132