DEPENDENCE OF MEASURED IONIZED CALCIUM ON PROTEIN-CONCENTRATION AS MEASURED BY 3 ION-SELECTIVE ELECTRODES

被引:6
作者
MCENROE, E
MCKENNA, MJ
DIAMOND, D
COVINGTON, AK
FREANEY, R
机构
[1] ST VINCENTS HOSP,METAB UNIT,CTR EDUC & RES,ELM PK,DUBLIN 4,IRELAND
[2] ST VINCENTS HOSP,DEPT ENDOCRINOL & DIABET,DUBLIN 4,IRELAND
[3] DUBLIN CITY UNIV,SCH CHEM SCI,DUBLIN,IRELAND
[4] ST MICHAELS HOSP,DEPT MED,DUBLIN,IRELAND
[5] UNIV NEWCASTLE UPON TYNE,DEPT CHEM,NEWCASTLE TYNE NE1 7RU,TYNE & WEAR,ENGLAND
关键词
REFERENCE METHOD; VENOUS PRESSURE; ALBUMIN; LIQUID JUNCTION;
D O I
10.1177/000456329202900413
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A positive effect of protein on the measurement of ionized calcium in serum by ion-selective electrodes (ISEs) has been previously reported and the present study confirms this finding. Ionized calcium in serum was measured in the presence of increasing protein concentrations induced by venous stasis in 17 healthy volunteer subjects. Ionized calcium was measured using two commercial analysers, a Radiometer ICA2 analyser and a Baker Analyte + 2 analyser, and a calcium cell devised by Covington for the calcium reference method (CRM). Both commercial analysers used charged ionophores and the CRM used a neutral carrier ionophore in the selective membrane. A small but significant rise in ionized calcium with increasing protein was measured on all analysers. Substitution of isotonic KCl for saturated KCl in the reference electrode of the CRM resulted in significantly reduced values for ionized calcium in paired serum samples when measured using the isotonic salt bridge. This study supports the premise that the positive effect of protein is related to the salt bridge concentration of the reference electrode rather than the ISE membrane composition.
引用
收藏
页码:443 / 449
页数:7
相关论文
共 14 条
[1]  
BOINK ABT, 1986, METHODOLOGY CLIN APP, V8, P39
[2]  
BUCKLEY BM, 1987, METHODOLOGY CLIN APP, V8, P299
[3]   CORRELATION BETWEEN SERUM IONIZED CALCIUM AND SERUM-ALBUMIN CONCENTRATIONS IN 2 HOSPITAL POPULATIONS [J].
BUTLER, SJ ;
PAYNE, RB ;
GUNN, IR ;
BURNS, J ;
PATERSON, CR .
BRITISH MEDICAL JOURNAL, 1984, 289 (6450) :948-950
[4]  
COVINGTON AK, 1989, METHODOLOGY CLIN APP, V11, P163
[5]  
COVINGTON AK, 1988, METHODOLOGY CLIN APP, V10, P119
[6]  
Ferreira P, 1979, J Automat Chem, V1, P94, DOI 10.1155/S1463924679000298
[7]   DETERMINATION OF IONIZED CALCIUM BY ION-SELECTIVE ELECTRODE IS NOT INDEPENDENT OF ALBUMIN CONCENTRATION [J].
FREANEY, R ;
EGAN, T ;
MCKENNA, MJ ;
DOOLIN, MC ;
MULDOWNEY, FP .
CLINICA CHIMICA ACTA, 1986, 158 (02) :129-137
[9]   PROTEIN INTERFERENCE WITH ION-SELECTIVE ELECTRODE MEASUREMENT DEPENDS ON REFERENCE ELECTRODE COMPOSITION AND DESIGN [J].
PAYNE, RB ;
BUCKLEY, BM ;
RAWSON, KM .
ANNALS OF CLINICAL BIOCHEMISTRY, 1991, 28 :68-72
[10]   EFFECTS OF IONIC-STRENGTH ON IONIZED CALCIUM MEASURED USING REFERENCE ELECTRODE LIQUID JUNCTIONS OF VARYING DESIGN AND COMPOSITION [J].
PAYNE, RB .
ANNALS OF CLINICAL BIOCHEMISTRY, 1991, 28 :235-239