COMPUTER-CONTROLLED STEREOTAXIC RADIOTHERAPY SYSTEM

被引:11
|
作者
HOUDEK, PV
SCHWADE, JG
SERAGO, CF
LANDY, HJ
PISCIOTTA, V
WU, XD
MARKOE, AM
LEWIN, AA
ABITBOL, AA
BUJNOSKI, JL
MARIENBERG, ES
FIEDLER, JA
GINSBERG, MS
机构
[1] UNIV MIAMI,SCH MED,DEPT NEUROSURG,MIAMI,FL 33152
[2] UNIV MIAMI,SCH MED,DEPT MED INSTRUMENTAT,MIAMI,FL 33152
关键词
STEREOTAXIC RADIOTHERAPY; COMPUTER-CONTROLLED STEREOTAXIC GUIDE; INTEGRATION OF DIAGNOSTIC AND THERAPEUTIC PROCEDURES; CT SIMULATOR;
D O I
10.1016/0360-3016(92)90997-V
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A computer-controlled stereotaxic radiotherapy system based on a low-frequency magnetic field technology integrated with a single fixation point stereotaxic guide has been designed and instituted. The magnetic field, generated in space by a special field source located in the accelerator gantry, is digitized in real time by a field sensor that is a six degree-of-freedom measurement device. As this sensor is an integral part of the patient stereotaxic halo, the patient position (x, y, z) and orientation (azimuth, elevation, roll) within the accelerator frame of reference are always known. Six parameters - three coordinates and three Euler space angles - are continuously transmitted to a computer where they are analyzed and compared with the stereotaxic parameters of the target point. Hence, the system facilitates rapid and accurate patient set-up for stereotaxic treatment as well as monitoring of patient during the subsequent irradiation session. The stereotaxic system has been developed to promote the integration of diagnostic and therapeutic procedures, with the specific aim of integrating CT and/or MR aided tumor localization and long term (4- to 7-week) fractionated radiotherapy of small intracranial and ocular lesions.
引用
收藏
页码:175 / 180
页数:6
相关论文
共 50 条
  • [31] VERSATILE SYSTEM FOR COMPUTER-CONTROLLED ASSEMBLY
    AMBLER, AP
    BARROW, HG
    BROWN, CM
    BURSTALL, RM
    POPPLESTONE, RJ
    ARTIFICIAL INTELLIGENCE, 1975, 6 (02) : 129 - 156
  • [32] COMPUTER-CONTROLLED LEARNING IN A SIMPLE SYSTEM
    TOSNEY, T
    HOYLE, G
    PROCEEDINGS OF THE ROYAL SOCIETY SERIES B-BIOLOGICAL SCIENCES, 1977, 195 (1120): : 365 - 393
  • [33] Design of a computer-controlled multileaf collimator for advanced electron radiotherapy
    Gauer, T.
    Albers, D.
    Cremers, F.
    Harmansa, R.
    Pellegrini, R.
    Schmidt, R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (23): : 5987 - 6003
  • [34] COMPUTER-CONTROLLED ADAPTIVE ANTENNA SYSTEM.
    Fetterolf, Peter C.
    Price, Kent M.
    Proceedings - IEEE Military Communications Conference, 1985, : 530 - 533
  • [35] A COMPUTER-CONTROLLED EXPERT SYSTEM FOR ORTHODONTIC ADVICE
    SIMSWILLIAMS, JH
    BROWN, ID
    MATTHEWMAN, A
    STEPHENS, CD
    BRITISH DENTAL JOURNAL, 1987, 163 (05) : 161 - 166
  • [36] THE TOOLING OF AN AUTOMATED COMPUTER-CONTROLLED MANUFACTURING SYSTEM
    SOLOUKHINA, NM
    SOVIET ENGINEERING RESEARCH, 1982, 2 (07): : 56 - 57
  • [37] A computer-controlled system for studying the thermodesorption of deuterium
    Kalachev, A.M.
    Kurnaev, V.A.
    Levchuk, D.V.
    Pribory i Tekhnika Eksperimenta, 2003, 46 (02): : 151 - 155
  • [38] SURVEILLANCE ASPECTS OF A COMPUTER-CONTROLLED TRAFFIC SYSTEM
    GORDON, RL
    DODGE, KW
    SCOTT, JE
    IEEE TRANSACTIONS ON VEHICULAR TECHNOLOGY, 1970, VT19 (01) : 90 - &
  • [39] COMPUTER-CONTROLLED PHOTO-ELECTROCHEMICAL SYSTEM
    HAAK, R
    TENCH, D
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1983, 130 (08) : C338 - C338
  • [40] Computer-controlled endoscopic performance assessment system
    G. B. Hanna
    T. Drew
    P. Clinch
    B. Hunter
    A. Cuschieri
    Surgical Endoscopy, 1998, 12 : 997 - 1000