AN EVALUATION OF JANSSON METHOD TO DECONVOLVE OVERLAPPED GAS-CHROMATOGRAPHIC PEAKS

被引:3
作者
CRILLY, PB
机构
[1] Hewlett-Packard Co, United States
来源
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS | 1988年 / 93卷 / 03期
关键词
Chemical Analysis--Equipment;
D O I
10.6028/jres.093.099
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The author presents the results of a comprehensive evaluation of Jansson's Method to resolve overlapped peaks that have been generated from two different sample mixtures. As a benchmark, these peaks were also resolved using shoulder quantitation and perpendicular drop techniques. As a second benchmark, these same sample mixtures were analyzed using an instrument with a relatively long column so that the peaks were fully resolved without using non-chemical peak resolving methods. Experimental results indicate that the quantitation accuracy of peaks resolved using Jansson's Method is similar to what can be obtained using an instrument with a relatively long column and is about an order of magnitude better than the graphical methods.
引用
收藏
页码:413 / 416
页数:4
相关论文
共 5 条
[1]  
ALTMAYER L, 1979, 89 HEWL TECHN PAP
[2]  
CRILLY P, 1987, THESIS NEW MEXICO ST
[3]  
Crilly P. B., 1987, J CHEMOM, V1, P79
[4]  
JANSSON PA, 1984, DECONVOLUTION APPLIC
[5]  
MIKKELSON L, 1971, 45 HEWL TECHN PAP