Patient specific parametric geometric modelling and finite element analysis of cementless hip prosthesis This paper proposes a framework for subject-specific cementless hip implant design and virtual assembly analysis of the instantiated stem with femur model using finite element method

被引:13
作者
George, Subin P. [1 ]
Kumar, Gurunathan Saravana [1 ]
机构
[1] Indian Inst Technol Madras, Dept Engn Design, Chennai, Tamil Nadu, India
关键词
hip prosthesis; custom design; feature based parametric CAD; virtual assembly analysis; finite element analysis;
D O I
10.1080/17452759.2012.755654
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This work proposes a framework for subject-specific cementless hip implant design and analysis. A set of pre-specified femoral features that can be used for custom implant design have been identified and extracted from the femur geometry reconstructed from computed tomography images. A parametric implant design is proposed based on the extracted femur features. Virtual assembly of the instantiated stem with femur model is done to check for the form fit. A finite element analysis of the resected femur with implant assembly is done for a known experimental loading condition to assess change in the stress-strain state of the femur and the stress shielding effect due to insertion of the implant. A commercially used modular implant design is chosen for the comparison studies in terms of form fit as well as stress shielding effect. The study shows that the proposed custom implant fairs well in both these aspects.
引用
收藏
页码:65 / 83
页数:19
相关论文
共 42 条
[1]  
Anderson J., 1991, MATER DESIGN, V12, P103
[2]   Comparison of isotropic and orthotropic material property assignments on femoral finite element models under two loading conditions [J].
Baca, Vaclav ;
Horak, Zdenek .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (08) :935-935
[3]   Preclinical cost analysis of orthopaedic implants: a custom versus standard cementless femoral component for revision total hip arthroplasty [J].
Chang, PB ;
Robie, BH ;
Bartel, DL .
JOURNAL OF BIOMECHANICS, 1999, 32 (12) :1309-1318
[4]   A critical analysis of stress shielding evaluation of hip prostheses [J].
Cristofolini, L .
CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 1997, 25 (4-5) :409-483
[5]   The longitudinal Young's modulus of cortical bone in the midshaft of human femur and its correlation with CT scanning data [J].
Cuppone, M ;
Seedhom, BB ;
Berry, E ;
Ostell, AE .
CALCIFIED TISSUE INTERNATIONAL, 2004, 74 (03) :302-309
[6]  
Dorr LD, 2006, HIP ARTHROPLASTY MIN
[7]   MECHANOTRANSDUCTION AND THE FUNCTIONAL-RESPONSE OF BONE TO MECHANICAL STRAIN [J].
DUNCAN, RL ;
TURNER, CH .
CALCIFIED TISSUE INTERNATIONAL, 1995, 57 (05) :344-358
[8]   A contact model with ingrowth control for bone remodelling around cementless stems [J].
Fernandes, PR ;
Folgado, J ;
Jacobs, C ;
Pellegrini, V .
JOURNAL OF BIOMECHANICS, 2002, 35 (02) :167-176
[9]  
FROST HM, 1983, CLIN ORTHOP RELAT R, V200, P198
[10]  
Fuku H, 1994, Nihon Seikeigeka Gakkai Zasshi, V68, P763