THE PULSAIR-KEELER NONCONTACT TONOMETER IN SELF-TONOMETRY - PRELIMINARY-RESULTS

被引:13
作者
CARENINI, BB
BROGLIATTI, B
TONETTO, C
RENIS, E
机构
[1] Ophthalmological Clinic, University of Turin, Turin, 10122
关键词
SELF-TONOMETRY; NONCONTACT TONOMETER; INTRAOCULAR PRESSURE (IOP); GLAUCOMA; GOLDMANN TONOMETER;
D O I
10.1007/BF00917979
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The authors considered the possibility of using the Pulsair-Keeler non-contact tonometer in self-tonometry. For this reason, 45 patients have been trained to use the instrument and, after a reasonable trial period. were able to self-measure their IOP. The results have been compared to the tonometric values measured with Goldmann tonometer to evaluate statistically the limits and the dependability of this method of measurement. The results shows that self-tonometry with the Pulsair-Keeler tonometer can be used in monitoring glaucomatous patients at home.
引用
收藏
页码:295 / 297
页数:3
相关论文
共 9 条
[1]  
DIBARI A, 1989, B OCUL, V68, P219
[2]  
DIBARI A, 1988, B OCUL, V62, P211
[3]  
DIBARI A, 1988, B OCUL, V67, P157
[4]  
DRAEGER J, 1991, 4 EUOP COURS RES OPH, P23
[5]   NONCONTACT APPLANATION TONOMETER - DESCRIPTION AND CLINICAL EVALUATION [J].
FORBES, M ;
PICO, G ;
GROLMAN, B .
ARCHIVES OF OPHTHALMOLOGY, 1974, 91 (02) :134-140
[6]   HOME TONOMETRY [J].
JENSEN, AD ;
MAUMENEE, AE .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1973, 76 (06) :929-932
[7]  
WILENSKY JT, 1987, ARCH OPHTHALMOL-CHIC, V105, P1072
[8]  
ZEIMER RC, 1986, ARCH OPHTHALMOL-CHIC, V104, P49
[9]  
ZEIMER RC, 1983, ARCH OPHTHALMOL-CHIC, V101, P1971