FOCAL SPACING AND NEAR-FIELD HEATING DURING PULSED HIGH-TEMPERATURE ULTRASOUND THERAPY

被引:125
作者
DAMIANOU, C [1 ]
HYNYNEN, K [1 ]
机构
[1] UNIV ARIZONA,ARIZONA HLTH SCI CTR,DEPT RADIAT ONCOL,TUCSON,AZ 85724
关键词
ULTRASOUND; TEMPERATURE; THERMAL DOSE; HYPERTHERMIA; SURGERY;
D O I
10.1016/0301-5629(93)90094-5
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
It has been proposed that high temperature short duration hyperthermia treatment would be perfusion insensitive and thus, significantly improved thermal exposure uniformity could be achieved. This study investigates the execution of such a treatment, which utilizes single spherically curved transducer and multiple sonications to cover the complete target volume. The spacing of neighboring pulses as a function of the transducer characteristics was studied utilizing computer simulations. In addition, the temperature elevation in front of the focal zone during multiple sonications was evaluated. It was found that significant delays (20 s or longer) between the sonications must be introduced in order to avoid unwanted tissue damage in front of the focal zone. In addition, decreasing the pulse duration and F-number reduced the temperature build-up in front of the focus. The results were verified in vivo in dog's thigh muscle. This study is important not only for hyperthermia but also for ultrasound surgery, and indicates that each sonication system must be carefully evaluated for potential thermal damage outside of the target volume prior to implementation in therapy.
引用
收藏
页码:777 / 787
页数:11
相关论文
共 27 条
[1]   AN ANNULAR FOCUS ULTRASONIC LENS FOR LOCAL HYPERTHERMIA TREATMENT OF SMALL TUMORS [J].
BEARD, RE ;
MAGIN, RL ;
FRIZZELL, LA ;
CAIN, CA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (02) :177-184
[2]   EFFECTS OF PHYSICAL PARAMETERS ON HIGH-TEMPERATURE ULTRASOUND HYPERTHERMIA [J].
BILLARD, BE ;
HYNYNEN, K ;
ROEMER, RB .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (04) :409-420
[3]  
BRITT RH, 1984, PROG EXP TUMOR RES, V28, P232
[4]   CONCENTRIC-RING AND SECTOR-VORTEX PHASED-ARRAY APPLICATORS FOR ULTRASOUND HYPERTHERMIA [J].
CAIN, CA ;
UMEMURA, S .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1986, 34 (05) :542-551
[5]   ULTRASONIC VELOCITY AND ATTENUATION IN MAMMALIAN-TISSUES [J].
CHIVERS, RC ;
PARRY, RJ .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1978, 63 (03) :940-953
[6]  
DAVIS B, 1989, WIN ANN M AM SOC MEC, V12, P51
[7]   THE EFFECTS OF TISSUE HETEROGENEITIES AND LARGE BLOOD-VESSELS ON THE THERMAL EXPOSURE INDUCED BY SHORT HIGH-POWER ULTRASOUND PULSES [J].
DORR, LN ;
HYNYNEN, K .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (01) :45-59
[8]   A SPHERICAL-SECTION ULTRASOUND PHASED-ARRAY APPLICATOR FOR DEEP LOCALIZED HYPERTHERMIA [J].
EBBINI, ES ;
CAIN, CA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (07) :634-643
[9]   THE EFFECT OF WAVE REFLECTION AND REFRACTION AT SOFT-TISSUE INTERFACES DURING ULTRASOUND HYPERTHERMIA TREATMENTS [J].
FAN, XB ;
HYNYNEN, K .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1992, 91 (03) :1727-1736
[10]  
FRY FJ, 1978, ULTRASOUND ITS APP 2, P689