Improved peripheral dose calculation accuracy for a small MLC field brought by the latest commercial treatment planning system

被引:5
作者
Chow, J. C. L. [1 ,2 ]
Grigorov, G. N. [3 ]
Jiang, R. [2 ,3 ]
机构
[1] Univ Hlth Network, Princess Margaret Hosp, Toronto, ON, Canada
[2] Univ Waterloo, Phys Dept, Waterloo, ON, Canada
[3] Grand River Hosp, Med Phys Dept, Toronto, ON, Canada
关键词
Intensity modulated radiotherapy; multi-leaf collimator; treatment planning software commissioning and dosimetry;
D O I
10.1017/S1460396906000161
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A recently released Pinnacle treatment planning system software, v7.4f includes some new physics features such as modeling of the rounded multi-leaf collimator (MLC) leaf ends and the tongue-and-groove structure between leaves. In this study, the above physics modeling improvements were verified by comparing the peripheral dose profiles for the small MLC fields calculated by the new Pinnacle v7.4f and the old Pinnacle v6.2b with those obtained from measurements experimentally. Three test MLC fields with different jaw sizes were prepared, and specific dose profiles (along cross-line, in-line and diagonal axis) at different depths were measured using a Varian 21 EX linear accelerator with 120-leaf Millennium MLC, big scanning water tank and photon diode. Estimated dose profiles for the test fields were calculated using Pinnacle v6.2b and v7.4f. By comparing the measured and calculated results, it was found that both v6.2b and v7.4f performed well in calculating the cross-line (along the gap between the longitudinal lengths of two leaves) and diagonal axis dose profiles at different depths. However, v7.4f gave calculated dose values closer to the measured field for in-line (gap between junctions of two rounded leaf ends) axis dose profiles at different depths. For the shape of the profile along the in-line axis, v7.4f calculated a flat "platform" dose profile of about 34.3% (inter-bank leakage) at depth dmax beyond the MLC field edge using a clinical dose grid size of 0.4 x 0.4 x 0.4 cm(3), compared to the "zigzag" dose profile varying between 35.4% and 42.1% measured using the water tank and diode. However, both Pinnacle v6.2b and v7.4f calculated the percentage depth dose for the test fields well compared to the measurements.
引用
收藏
页码:121 / 128
页数:8
相关论文
共 19 条
[1]   A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy [J].
Arnfield, MR ;
Siebers, JV ;
Kim, JO ;
Wu, QW ;
Keall, PJ ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (10) :2231-2241
[2]   A variable fluence step clustering and segmentation algorithm for step and shoot IMRT [J].
Bär, W ;
Alber, M ;
Nüsslin, F .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (07) :1997-2007
[3]   Geometric analysis of light-field position of a multileaf collimator with curved ends [J].
Boyer, AL ;
Li, SD .
MEDICAL PHYSICS, 1997, 24 (05) :757-762
[4]   Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system [J].
Cadman, P ;
Bassalow, R ;
Sidhu, NPS ;
Ibbott, G ;
Nelson, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (16) :3001-3010
[5]   Validation of physics improvements for IMRT with a commercial treatment-planning system [J].
Cadman, Patrick ;
McNutt, Todd ;
Bzdusek, Karl .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2005, 6 (02) :74-86
[6]   Dosimetric effects on the penumbra region of irregular multi-leaf collimated fields [J].
Chow, JCL ;
Wettlaufer, B ;
Jiang, RQ .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (03) :N31-N38
[7]   Dosimetric effect of collimating jaws for small multileaf collimated fields [J].
Chow, JCL ;
Seguin, M ;
Alexander, A .
MEDICAL PHYSICS, 2005, 32 (03) :759-765
[8]   Beam characteristics of a retrofitted double-focused multileaf collimator [J].
Das, IJ ;
Desobry, GE ;
McNeeley, SW ;
Cheng, EC ;
Schultheiss, TE .
MEDICAL PHYSICS, 1998, 25 (09) :1676-1684
[9]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[10]  
Galvin J., 2001, BASIC APPL MULTILEAF